P551Permanent cardiac implantable device damage during direct photon exposure for oncologic radiotherapy: a multicentre, in-vitro observation. (18th June 2020)
- Record Type:
- Journal Article
- Title:
- P551Permanent cardiac implantable device damage during direct photon exposure for oncologic radiotherapy: a multicentre, in-vitro observation. (18th June 2020)
- Main Title:
- P551Permanent cardiac implantable device damage during direct photon exposure for oncologic radiotherapy: a multicentre, in-vitro observation
- Authors:
- Di Girolamo, E
Faustino, M
Furia, N
Appignani, M
Marini, M
De Filippo, P
Leidi, C
Delana, A
Barbareschi, A
Andreoli, S
Genovesi, D
Falco, M D - Abstract:
- Abstract: Funding Acknowledgements: None Backgroung. Direct photon exposure of cardiac implantable devices (CIEDs), both pacemakers (PMs) or implantable cardioverter defibrillators (ICDs), during oncologic radiotherapy (RT) courses may transiently or permanently affect normal device function. Purpose: To evaluate CIED damage by direct exposure to doses up to 10 Gy in oncologic RT, commonly considered unsafe or even potentially harmful, 206 CIEDs (143 PMs and 63 ICDs) from three different centres, with at least 4 months to Elective Replacement Indicator (E.R.I.) were observed. Methods. All CIEDs had a baseline telemetry interrogation. Single chamber devices were programmed in the VVI/40 mode and dual or triple chamber ones were programmed in the DDD/40 mode. Rate adaptive function was disabled. In ICDs, antitachycardia therapies were disabled with the ventricular tachycardia/fibrillation window left enabled. A centering Computed Tomography was performed to build the corresponding treatment plan and CIEDs were blinded randomized to receive either 2, 5 or 10 Gy (direct exposure) by a 6 MV linear accelerator in a home-made water phantom. An in-vivo dosimetry randomly assessed the effective dose received by the CIEDs. All CIEDs were interrogated immediately after exposure and monthly during a three-month follow-up. Results. Immediately after photon exposure, no changes in device setting or software errors were observed in 205 CIEDs (99·5%). Reset to emergency back-up mode wasAbstract: Funding Acknowledgements: None Backgroung. Direct photon exposure of cardiac implantable devices (CIEDs), both pacemakers (PMs) or implantable cardioverter defibrillators (ICDs), during oncologic radiotherapy (RT) courses may transiently or permanently affect normal device function. Purpose: To evaluate CIED damage by direct exposure to doses up to 10 Gy in oncologic RT, commonly considered unsafe or even potentially harmful, 206 CIEDs (143 PMs and 63 ICDs) from three different centres, with at least 4 months to Elective Replacement Indicator (E.R.I.) were observed. Methods. All CIEDs had a baseline telemetry interrogation. Single chamber devices were programmed in the VVI/40 mode and dual or triple chamber ones were programmed in the DDD/40 mode. Rate adaptive function was disabled. In ICDs, antitachycardia therapies were disabled with the ventricular tachycardia/fibrillation window left enabled. A centering Computed Tomography was performed to build the corresponding treatment plan and CIEDs were blinded randomized to receive either 2, 5 or 10 Gy (direct exposure) by a 6 MV linear accelerator in a home-made water phantom. An in-vivo dosimetry randomly assessed the effective dose received by the CIEDs. All CIEDs were interrogated immediately after exposure and monthly during a three-month follow-up. Results. Immediately after photon exposure, no changes in device setting or software errors were observed in 205 CIEDs (99·5%). Reset to emergency back-up mode was observed in a PM (0·49% overall; 0·7% among PMs). Seven PMs reached the E.R.I immediately after exposure (3·4% overall; 4·9% among PMs). Sixteen ICDs (7·8% overall; 25·4% among ICDs) reported multiple ventricular tachycardia/fibrillation detections stored in the device memory. During follow-up, a non-reprogrammable software reset (emergency backup VVI/65 mode) was observed in one PM after a single dose of 2 Gy (0·49% overall; 0.7% among PMs), whereas an abnormal battery drain was observed in 6 PMs (2.9% overall; 4.2% among PMs). No battery issues were observed in ICDs. All reported events occurred regardless of either 2, 5, or 10 Gy direct exposure. Malfunctions were observed in only older CIEDs. Conclusions. Recent CIEDs have shown to be safe during oncologic RT, withstanding direct exposure up to 10 Gy, commonly considered not recommended or even unsafe by manufacturers statements and clinical guidelines. Malfunctions occurred solely in older devices. … (more)
- Is Part Of:
- Europace. Volume 22(2020)Supplement 1
- Journal:
- Europace
- Issue:
- Volume 22(2020)Supplement 1
- Issue Display:
- Volume 22, Issue 1 (2020)
- Year:
- 2020
- Volume:
- 22
- Issue:
- 1
- Issue Sort Value:
- 2020-0022-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-06-18
- Subjects:
- Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euaa162.113 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15327.xml