A Comparison of Body Composition Measurements Between Bioelectrical Impedance Analysis (InBody 570) and Air Displacement Plethysmography (BOD POD®). (29th May 2020)
- Record Type:
- Journal Article
- Title:
- A Comparison of Body Composition Measurements Between Bioelectrical Impedance Analysis (InBody 570) and Air Displacement Plethysmography (BOD POD®). (29th May 2020)
- Main Title:
- A Comparison of Body Composition Measurements Between Bioelectrical Impedance Analysis (InBody 570) and Air Displacement Plethysmography (BOD POD®)
- Authors:
- Sirirat, Rawiwan
Heskey, Celine
Wilson, Christine
Bitok, Edward
Jones, Julie
Clarke, Abigail
Sabaté, Joan - Abstract:
- Abstract: Objectives: The accurate measurement of body composition is important in both research and clinical practice. The aim of this study was to compare the InBody relative to the BOD POD®. The latter is widely recognized as one of the most accurate methods to measure human body composition. Methods: In the context of a clinical trial of 35 free-living non-athletic individuals [80% F, ages 40–69 years, BMI 25–34 kg/m2], we compared body composition measurements utilizing Bioelectrical Impedance Analysis (BIA) and Air Displacement Plethysmography (ADP). ADP was conducted in a BOD POD® (Cosmed USA Inc., Concord, CA, USA) and BIA measured using InBody 570 (In Body, Cerritos, CA, USA). Body measurements included total body weight, fat mass and fat-free mass which were obtained in kilograms following manufacturer instructions. Spearman's rank (r s ) and Pearson correlations (r) were used to evaluate the agreement between the two instruments. Results: The BOD POD® and InBody measurements were strongly correlated. Correlation was strongest for total body weight (r s (35) = .99, P < .0001), followed by fat mass (r (35) = .93, P < .0001). The lowest correlation was observed for fat-free mass (r s (35) = .79, P < .0001). Conclusions: The InBody 570 is reliable and compares favorably to the BOD POD®. Hence, it can be used in clinical settings and epidemiological studies as a practical and relatively inexpensive alternative to the BodPod and dual-energy x-ray absorptiometryAbstract: Objectives: The accurate measurement of body composition is important in both research and clinical practice. The aim of this study was to compare the InBody relative to the BOD POD®. The latter is widely recognized as one of the most accurate methods to measure human body composition. Methods: In the context of a clinical trial of 35 free-living non-athletic individuals [80% F, ages 40–69 years, BMI 25–34 kg/m2], we compared body composition measurements utilizing Bioelectrical Impedance Analysis (BIA) and Air Displacement Plethysmography (ADP). ADP was conducted in a BOD POD® (Cosmed USA Inc., Concord, CA, USA) and BIA measured using InBody 570 (In Body, Cerritos, CA, USA). Body measurements included total body weight, fat mass and fat-free mass which were obtained in kilograms following manufacturer instructions. Spearman's rank (r s ) and Pearson correlations (r) were used to evaluate the agreement between the two instruments. Results: The BOD POD® and InBody measurements were strongly correlated. Correlation was strongest for total body weight (r s (35) = .99, P < .0001), followed by fat mass (r (35) = .93, P < .0001). The lowest correlation was observed for fat-free mass (r s (35) = .79, P < .0001). Conclusions: The InBody 570 is reliable and compares favorably to the BOD POD®. Hence, it can be used in clinical settings and epidemiological studies as a practical and relatively inexpensive alternative to the BodPod and dual-energy x-ray absorptiometry (DEXA). Funding Sources: Nutrition Research Center, School of Public Health, Loma Linda University, Loma Linda, CA, USA. … (more)
- Is Part Of:
- Current developments in nutrition. Volume 4(2020)Supplement 2
- Journal:
- Current developments in nutrition
- Issue:
- Volume 4(2020)Supplement 2
- Issue Display:
- Volume 4, Issue 2 (2020)
- Year:
- 2020
- Volume:
- 4
- Issue:
- 2
- Issue Sort Value:
- 2020-0004-0002-0000
- Page Start:
- 1689
- Page End:
- 1689
- Publication Date:
- 2020-05-29
- Subjects:
- Nutrition -- Periodicals
Nutritional Physiological Phenomena
Nutrition
Periodicals
Periodicals
Fulltext
Internet Resources
Periodicals
612.3 - Journal URLs:
- https://academic.oup.com/cdn ↗
https://www.sciencedirect.com/journal/current-developments-in-nutrition ↗
https://cdn.nutrition.org/ ↗
http://www.oxfordjournals.org/ ↗ - DOI:
- 10.1093/cdn/nzaa063_087 ↗
- Languages:
- English
- ISSNs:
- 2475-2991
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15319.xml