IFCC Working Group Recommendations for Assessing Commutability Part 3: Using the Calibration Effectiveness of a Reference Material. (1st March 2018)
- Record Type:
- Journal Article
- Title:
- IFCC Working Group Recommendations for Assessing Commutability Part 3: Using the Calibration Effectiveness of a Reference Material. (1st March 2018)
- Main Title:
- IFCC Working Group Recommendations for Assessing Commutability Part 3: Using the Calibration Effectiveness of a Reference Material
- Authors:
- Budd, Jeffrey R
Weykamp, Cas
Rej, Robert
MacKenzie, Finlay
Ceriotti, Ferruccio
Greenberg, Neil
Camara, Johanna E
Schimmel, Heinz
Vesper, Hubert W
Keller, Thomas
Delatour, Vincent
Panteghini, Mauro
Burns, Chris
Miller, W Greg - Abstract:
- Abstract: A process is described to assess the commutability of a reference material (RM) intended for use as a calibrator based on its ability to fulfill its intended use in a calibration traceability scheme to produce equivalent clinical sample (CS) results among different measurement procedures (MPs) for the same measurand. Three sources of systematic error are elucidated in the context of creating the calibration model for translating MP signals to measurand amounts: calibration fit, calibrator level trueness, and commutability. An example set of 40 CS results from 7 MPs is used to illustrate estimation of bias and variability for each MP. The candidate RM is then used to recalibrate each MP, and its effectiveness in reducing the systematic error among the MPs within an acceptable level of equivalence based on medical requirements confirms its commutability for those MPs. The RM is declared noncommutable for MPs for which, after recalibration, the CS results do not agree with those from other MPs. When a lack of agreement is found, other potential causes, including lack of calibration fit, should be investigated before concluding the RM is noncommutable. The RM is considered fit for purpose for those MPs where commutability is demonstrated.
- Is Part Of:
- Clinical chemistry. Volume 64:Number 3(2018)
- Journal:
- Clinical chemistry
- Issue:
- Volume 64:Number 3(2018)
- Issue Display:
- Volume 64, Issue 3 (2018)
- Year:
- 2018
- Volume:
- 64
- Issue:
- 3
- Issue Sort Value:
- 2018-0064-0003-0000
- Page Start:
- 465
- Page End:
- 474
- Publication Date:
- 2018-03-01
- Subjects:
- Clinical chemistry -- Periodicals
Pharmaceutical chemistry -- Periodicals
Biochemistry -- Periodicals
Biochimie -- Périodiques
Diagnostics biologiques -- Périodiques
Biochemistry
Clinical chemistry
Pharmaceutical chemistry
Biochemistry
Laboratory Techniques and Procedures
Klinische chemie
Periodicals
616.075605 - Journal URLs:
- http://www.oxfordjournals.org/ ↗
https://academic.oup.com/clinchem ↗
http://catalog.hathitrust.org/api/volumes/oclc/1554929.html ↗
http://www.clinchem.org/ ↗ - DOI:
- 10.1373/clinchem.2017.277558 ↗
- Languages:
- English
- ISSNs:
- 0009-9147
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15303.xml