Modeling of Effect of Glucose Sensor Errors on Insulin Dosage and Glucose Bolus Computed by LOGIC-Insulin. (1st December 2014)
- Record Type:
- Journal Article
- Title:
- Modeling of Effect of Glucose Sensor Errors on Insulin Dosage and Glucose Bolus Computed by LOGIC-Insulin. (1st December 2014)
- Main Title:
- Modeling of Effect of Glucose Sensor Errors on Insulin Dosage and Glucose Bolus Computed by LOGIC-Insulin
- Authors:
- Van Herpe, Tom
De Moor, Bart
Van den Berghe, Greet
Mesotten, Dieter - Abstract:
- Abstract: BACKGROUND: Effective and safe glycemic control in critically ill patients requires accurate glucose sensors and adequate insulin dosage calculators. The LOGIC-Insulin calculator for glycemic control has recently been validated in the LOGIC-1 randomized controlled trial. In this study, we aimed to determine the allowable error for intermittent and continuous glucose sensors, on the basis of the LOGIC-Insulin calculator. METHODS: A gaussian simulation model with a varying bias (0%–20%) and CV (−20% to +20%) simulated blood glucose values from the LOGIC-1 study (n = 149 patients) in 10 Monte Carlo steps. A clinical error grid system was developed to compare the simulated LOGIC-Insulin–directed intervention with the nominal intervention (0% bias, 0% CV). The severity of error measuring the clinical effect of the simulated LOGIC-Insulin intervention was graded as type B, C, and D errors. Type D errors were classified as acutely life-threatening (0% probability preferred). RESULTS: The probability of all types of errors was lower for continuous sensors compared with intermittent sensors. The maximum total error (TE), defined as the first TE introducing a type B/C/D error, was similar for both sensor types. To avoid type D errors, TEs <15.7% for intermittent sensors and <17.8% for continuous sensors were required. Mean absolute relative difference thresholds for type C errors were 7.1% for intermittent and 11.0% for continuous sensors. CONCLUSIONS: Continuous sensors hadAbstract: BACKGROUND: Effective and safe glycemic control in critically ill patients requires accurate glucose sensors and adequate insulin dosage calculators. The LOGIC-Insulin calculator for glycemic control has recently been validated in the LOGIC-1 randomized controlled trial. In this study, we aimed to determine the allowable error for intermittent and continuous glucose sensors, on the basis of the LOGIC-Insulin calculator. METHODS: A gaussian simulation model with a varying bias (0%–20%) and CV (−20% to +20%) simulated blood glucose values from the LOGIC-1 study (n = 149 patients) in 10 Monte Carlo steps. A clinical error grid system was developed to compare the simulated LOGIC-Insulin–directed intervention with the nominal intervention (0% bias, 0% CV). The severity of error measuring the clinical effect of the simulated LOGIC-Insulin intervention was graded as type B, C, and D errors. Type D errors were classified as acutely life-threatening (0% probability preferred). RESULTS: The probability of all types of errors was lower for continuous sensors compared with intermittent sensors. The maximum total error (TE), defined as the first TE introducing a type B/C/D error, was similar for both sensor types. To avoid type D errors, TEs <15.7% for intermittent sensors and <17.8% for continuous sensors were required. Mean absolute relative difference thresholds for type C errors were 7.1% for intermittent and 11.0% for continuous sensors. CONCLUSIONS: Continuous sensors had a lower probability for clinical errors than intermittent sensors at the same accuracy level. These simulations demonstrated the suitability of the LOGIC-Insulin control system for use with continuous, as well as intermittent, sensors. … (more)
- Is Part Of:
- Clinical chemistry. Volume 60:Number 12(2014)
- Journal:
- Clinical chemistry
- Issue:
- Volume 60:Number 12(2014)
- Issue Display:
- Volume 60, Issue 12 (2014)
- Year:
- 2014
- Volume:
- 60
- Issue:
- 12
- Issue Sort Value:
- 2014-0060-0012-0000
- Page Start:
- 1510
- Page End:
- 1518
- Publication Date:
- 2014-12-01
- Subjects:
- Clinical chemistry -- Periodicals
Pharmaceutical chemistry -- Periodicals
Biochemistry -- Periodicals
Biochimie -- Périodiques
Diagnostics biologiques -- Périodiques
Biochemistry
Clinical chemistry
Pharmaceutical chemistry
Biochemistry
Laboratory Techniques and Procedures
Klinische chemie
Periodicals
616.075605 - Journal URLs:
- http://www.oxfordjournals.org/ ↗
https://academic.oup.com/clinchem ↗
http://catalog.hathitrust.org/api/volumes/oclc/1554929.html ↗
http://www.clinchem.org/ ↗ - DOI:
- 10.1373/clinchem.2014.227017 ↗
- Languages:
- English
- ISSNs:
- 0009-9147
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15280.xml