Implantation technique and optimal subcutaneous defibrillator chest position: a PRAETORIAN score-based study. Issue 12 (29th October 2020)
- Record Type:
- Journal Article
- Title:
- Implantation technique and optimal subcutaneous defibrillator chest position: a PRAETORIAN score-based study. Issue 12 (29th October 2020)
- Main Title:
- Implantation technique and optimal subcutaneous defibrillator chest position: a PRAETORIAN score-based study
- Authors:
- Francia, Pietro
Biffi, Mauro
Adduci, Carmen
Ottaviano, Luca
Migliore, Federico
De Bonis, Silvana
Dello Russo, Antonio
De Filippo, Paolo
Viani, Stefano
Bongiorni, Maria Grazia
Caravati, Fabrizio
Lavalle, Carlo
Landolina, Maurizio Eugenio
Pisanò, Ennio
Giorgi, Davide
Lovecchio, Mariolina
Valsecchi, Sergio
Diemberger, Igor - Abstract:
- Abstract: Aims: The traditional technique for subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation involves three incisions and a subcutaneous pocket. Recently, a two-incision and intermuscular (IM) technique has been adopted. The PRAETORIAN score is a chest radiograph-based tool that predicts S-ICD conversion testing. We assessed whether the S-ICD implantation technique affects optimal position of the defibrillation system according to the PRAETORIAN score. Methods and results: We analysed consecutive patients undergoing S-ICD implantation. The χ 2 test and regression analysis were used to determine the association between the PRAETORIAN score and implantation technique. Two hundred and thirteen patients were enrolled. The S-ICD generator was positioned in an IM pocket in 174 patients (81.7%) and the two-incision approach was adopted in 199 (93.4%). According to the PRAETORIAN score, the risk of conversion failure was classified as low in 198 patients (93.0%), intermediate in 13 (6.1%), and high in 2 (0.9%). Patients undergoing the two-incision and IM technique were more likely to have a low (<90) PRAETORIAN score than those undergoing the three-incision and subcutaneous technique (two-incision: 94.0% vs. three-incision: 78.6%; P = 0.004 and IM: 96.0% vs. subcutaneous: 79.5%; P = 0.001). Intermuscular plus two-incision technique was associated with a low-risk PRAETORIAN score (hazard ratio 3.76; 95% confidence interval 1.01–14.02; P = 0.04). ShockAbstract: Aims: The traditional technique for subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation involves three incisions and a subcutaneous pocket. Recently, a two-incision and intermuscular (IM) technique has been adopted. The PRAETORIAN score is a chest radiograph-based tool that predicts S-ICD conversion testing. We assessed whether the S-ICD implantation technique affects optimal position of the defibrillation system according to the PRAETORIAN score. Methods and results: We analysed consecutive patients undergoing S-ICD implantation. The χ 2 test and regression analysis were used to determine the association between the PRAETORIAN score and implantation technique. Two hundred and thirteen patients were enrolled. The S-ICD generator was positioned in an IM pocket in 174 patients (81.7%) and the two-incision approach was adopted in 199 (93.4%). According to the PRAETORIAN score, the risk of conversion failure was classified as low in 198 patients (93.0%), intermediate in 13 (6.1%), and high in 2 (0.9%). Patients undergoing the two-incision and IM technique were more likely to have a low (<90) PRAETORIAN score than those undergoing the three-incision and subcutaneous technique (two-incision: 94.0% vs. three-incision: 78.6%; P = 0.004 and IM: 96.0% vs. subcutaneous: 79.5%; P = 0.001). Intermuscular plus two-incision technique was associated with a low-risk PRAETORIAN score (hazard ratio 3.76; 95% confidence interval 1.01–14.02; P = 0.04). Shock impedance was lower in PRAETORIAN low-risk patients than in intermediate-/high-risk categories (66 vs. 96 Ohm; P = 0.001). The PRAETORIAN score did not predict shock failure at 65 J. Conclusion: In this cohort of S-ICD recipients, combining the two-incision technique and IM generator implantation yielded the lowest PRAETORIAN score values, indicating optimal defibrillation system position. Clinical trial registration: http://clinicaltrials.gov/ Identifier: NCT02275637. … (more)
- Is Part Of:
- Europace. Volume 22:Issue 12(2020)
- Journal:
- Europace
- Issue:
- Volume 22:Issue 12(2020)
- Issue Display:
- Volume 22, Issue 12 (2020)
- Year:
- 2020
- Volume:
- 22
- Issue:
- 12
- Issue Sort Value:
- 2020-0022-0012-0000
- Page Start:
- 1822
- Page End:
- 1829
- Publication Date:
- 2020-10-29
- Subjects:
- Sudden death -- Implantable cardioverter defibrillator -- Subcutaneous -- Defibrillation test -- Safety -- Conversion -- PRAETORIAN score
Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euaa231 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15256.xml