A multicentered evaluation of ablation at higher power guided by ablation index: Establishing ablation targets for pulmonary vein isolation. (6th January 2019)
- Record Type:
- Journal Article
- Title:
- A multicentered evaluation of ablation at higher power guided by ablation index: Establishing ablation targets for pulmonary vein isolation. (6th January 2019)
- Main Title:
- A multicentered evaluation of ablation at higher power guided by ablation index: Establishing ablation targets for pulmonary vein isolation
- Authors:
- Dhillon, Gurpreet
Ahsan, Syed
Honarbakhsh, Shohreh
Lim, Wei
Baca, Marco
Graham, Adam
Srinivasan, Neil
Sawhney, Vinit
Sporton, Simon
Schilling, Richard J.
Chow, Anthony
Ginks, Matthew
Sohal, Manav
Gallagher, Mark M.
Hunter, Ross J. - Abstract:
- Abstract: Background: Pulmonary vein isolation (PVI) using high power delivered by SmartTouch Surround Flow (STSF) catheters guided by ablation index (AI) was evaluated in a multicenter registry. Methods: Patients with paroxysmal AF underwent PVI with STSF catheters using 30 W on the posterior wall and 40 W elsewhere. AI targets were 350 posterior walls and 450 elsewhere. Procedures were compared with controls using conventionally irrigated contact force‐sensing catheters using conventional powers (25 W posterior wall and 30 W elsewhere) guided by force‐time integral (no agreed targets). The waiting period of 30 minutes was observed before adenosine administration to assess acute pulmonary vein (PV) reconnection. Results: One hundred patients from four centers were included: 50 patients in the high power ablation index (HPAI) group and 50 controls. Procedure time was 22% shorter in the HPAI group (156 [133.8‐179] vs 199 [178.5‐227] minutes; P < 0.001). Duration of the radiofrequency application was 37% shorter in the HPAI group (27.2 [21.5‐35.8] vs 43.2 [35.1‐52.1] minutes; P < 0.001). Acute PV reconnection was reduced (28 of 200 [14%] vs 48 of 200 [24%] veins; P = 0.015). Reconnection was predicted by a largest interlesion distance greater than 6 mm, a lesion with impedance drop less than 2.5 Ω, contact force less than 6 g, or less than 68% of the regional AI target (all P < 0.001). Freedom from atrial arrhythmia at 1 year off antiarrhythmic drugs after a singleAbstract: Background: Pulmonary vein isolation (PVI) using high power delivered by SmartTouch Surround Flow (STSF) catheters guided by ablation index (AI) was evaluated in a multicenter registry. Methods: Patients with paroxysmal AF underwent PVI with STSF catheters using 30 W on the posterior wall and 40 W elsewhere. AI targets were 350 posterior walls and 450 elsewhere. Procedures were compared with controls using conventionally irrigated contact force‐sensing catheters using conventional powers (25 W posterior wall and 30 W elsewhere) guided by force‐time integral (no agreed targets). The waiting period of 30 minutes was observed before adenosine administration to assess acute pulmonary vein (PV) reconnection. Results: One hundred patients from four centers were included: 50 patients in the high power ablation index (HPAI) group and 50 controls. Procedure time was 22% shorter in the HPAI group (156 [133.8‐179] vs 199 [178.5‐227] minutes; P < 0.001). Duration of the radiofrequency application was 37% shorter in the HPAI group (27.2 [21.5‐35.8] vs 43.2 [35.1‐52.1] minutes; P < 0.001). Acute PV reconnection was reduced (28 of 200 [14%] vs 48 of 200 [24%] veins; P = 0.015). Reconnection was predicted by a largest interlesion distance greater than 6 mm, a lesion with impedance drop less than 2.5 Ω, contact force less than 6 g, or less than 68% of the regional AI target (all P < 0.001). Freedom from atrial arrhythmia at 1 year off antiarrhythmic drugs after a single procedure was 78% in the HPAI group vs 64% in the control group ( P = 0.186). Conclusion: High‐powered ablation guided by AI was safe and led to shorter procedure times with reduced acute PV reconnection compared with conventional ablation. … (more)
- Is Part Of:
- Journal of cardiovascular electrophysiology. Volume 30:Number 3(2019)
- Journal:
- Journal of cardiovascular electrophysiology
- Issue:
- Volume 30:Number 3(2019)
- Issue Display:
- Volume 30, Issue 3 (2019)
- Year:
- 2019
- Volume:
- 30
- Issue:
- 3
- Issue Sort Value:
- 2019-0030-0003-0000
- Page Start:
- 357
- Page End:
- 365
- Publication Date:
- 2019-01-06
- Subjects:
- ablation index -- atrial fibrillation -- pulmonary vein isolation
Blood vessels -- Physiology -- Periodicals
Electrophysiology -- Periodicals
Heart -- Physiology -- Periodicals
612.1 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/jce.13813 ↗
- Languages:
- English
- ISSNs:
- 1045-3873
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 15229.xml