Commissioning and performance testing of the first prototype of AlignRT InBore™ a Halcyon™ and Ethos™-dedicated surface guided radiation therapy platform. (December 2020)
- Record Type:
- Journal Article
- Title:
- Commissioning and performance testing of the first prototype of AlignRT InBore™ a Halcyon™ and Ethos™-dedicated surface guided radiation therapy platform. (December 2020)
- Main Title:
- Commissioning and performance testing of the first prototype of AlignRT InBore™ a Halcyon™ and Ethos™-dedicated surface guided radiation therapy platform
- Authors:
- Nguyen, Daniel
Farah, Jad
Barbet, Nicolas
Khodri, Mustapha - Abstract:
- Highlights: AlingRT InBore™ is a novel SGRT solution for bore-based Linacs. Cameras caused no adverse effects on beam delivery or image quality. Sub-millimetric accuracy is achieved for inter and intrafraction motion management. Systems' stability and performance are not affected by Halcyon™ rotations. AlignRT InBore™ shall allow contactless DIBH treatments on Halcyon™. Abstract: Purpose: To commission and assess the performance of AlignRT InBore™, a Halcyon™ and Ethos™-dedicated Surface Guided Radiation Therapy (SGRT) platform which combines ceiling-mounted cameras for patient setup and bore-mounted cameras for in-bore tracking. Methods: To check the potential impact of InBore™ cameras on dose delivery, 16 SRS, H&N, breast and pelvis patients' quality assurance (QA) treatment plans were measured with/without AlignRT InBore™ and using ArcCHECK® and SRS MapCHECK®. Impact on image quality was determined using Catphan® 540 phantom and considering all available MV and CBCT protocols (head, breast, chest and pelvis). The stability, accuracy and overall performance of AlignRT InBore™ was assessed using an MV Cube and anthropomorphic phantoms. Results: Comparison of 2D dose distributions with/without AlignRT InBore™ showed no impact on treatment delivery for all 16 QA checks (p-value > 0.25). 2D and CBCT images showed no artefacts or change in the contrast-to-noise ratio, resolution and noise values measured with Catphan® 540. Anti-collision sensors were unaffected by theHighlights: AlingRT InBore™ is a novel SGRT solution for bore-based Linacs. Cameras caused no adverse effects on beam delivery or image quality. Sub-millimetric accuracy is achieved for inter and intrafraction motion management. Systems' stability and performance are not affected by Halcyon™ rotations. AlignRT InBore™ shall allow contactless DIBH treatments on Halcyon™. Abstract: Purpose: To commission and assess the performance of AlignRT InBore™, a Halcyon™ and Ethos™-dedicated Surface Guided Radiation Therapy (SGRT) platform which combines ceiling-mounted cameras for patient setup and bore-mounted cameras for in-bore tracking. Methods: To check the potential impact of InBore™ cameras on dose delivery, 16 SRS, H&N, breast and pelvis patients' quality assurance (QA) treatment plans were measured with/without AlignRT InBore™ and using ArcCHECK® and SRS MapCHECK®. Impact on image quality was determined using Catphan® 540 phantom and considering all available MV and CBCT protocols (head, breast, chest and pelvis). The stability, accuracy and overall performance of AlignRT InBore™ was assessed using an MV Cube and anthropomorphic phantoms. Results: Comparison of 2D dose distributions with/without AlignRT InBore™ showed no impact on treatment delivery for all 16 QA checks (p-value > 0.25). 2D and CBCT images showed no artefacts or change in the contrast-to-noise ratio, resolution and noise values measured with Catphan® 540. Anti-collision sensors were unaffected by the bore-mounted cameras. Additionally, AlignRT InBore™ cameras allowed for motion detection with sub-0.5 mm accuracy and sub-0.4 mm stability with surface coverage of >50 × 60 × 35 cc. Accurate transition (sub-0.3 mm) from virtual to treatment isocentres was achieved. Finally, Halcyon™ rotations during CBCT and beam delivery resulted in limited camera vibrations with translation uncertainty <0.5 mm in left-right and anterior-posterior directions and <0.1 mm in head-feet direction. Conclusion: AlignRT InBore™ provides SGRT setup and intrafraction monitoring capabilities with a performance comparable to standard SGRT solutions while having no adverse effect on Halcyon™. … (more)
- Is Part Of:
- Physica medica. Volume 80(2021)
- Journal:
- Physica medica
- Issue:
- Volume 80(2021)
- Issue Display:
- Volume 80, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 80
- Issue:
- 2021
- Issue Sort Value:
- 2021-0080-2021-0000
- Page Start:
- 159
- Page End:
- 166
- Publication Date:
- 2020-12
- Subjects:
- Surface guidance -- Setup verification -- Motion management -- Halcyon™
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2020.10.024 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15203.xml