A Novel Calibration Method for the Quantification of Respirable Particles in Mining Scenarios Using Fourier Transform Infrared Spectroscopy. Issue 3 (March 2021)
- Record Type:
- Journal Article
- Title:
- A Novel Calibration Method for the Quantification of Respirable Particles in Mining Scenarios Using Fourier Transform Infrared Spectroscopy. Issue 3 (March 2021)
- Main Title:
- A Novel Calibration Method for the Quantification of Respirable Particles in Mining Scenarios Using Fourier Transform Infrared Spectroscopy
- Authors:
- Stach, Robert
Barone, Teresa
Cauda, Emanuele
Mizaikoff, Boris - Abstract:
- The exposure of mining workers to crystalline particles, e.g., alpha quartz in respirable dust, is a ubiquitous global problem in occupational safety and health at surface and underground operations. The challenge of rapid in-field monitoring for direct assessment and adoption of intervention has not been solved satisfactorily to date, as conventional analytical methods such as X-ray diffraction and infrared spectroscopy require laboratory environments, complex system handling, tedious sample preparation, and are limited by, e.g., addressable particle size. A novel monitoring approach was developed for potential in-field application enabling the quantification of crystalline particles in the respirable regime based on transmission infrared spectroscopy. This on-site approach analyzes samples of dust in ambient air collected onto PVC filters using respirable dust sampling devices. In the present study, we demonstrate that portable Fourier transform infrared (FT-IR) spectroscopy in combination with multivariate data analysis provides a versatile tool for the identification and quantification of minerals in complex real-world matrices. Without further sample preparation, the loaded filters are immediately analyzed via transmission infrared spectroscopy, and the mineral amount is quantified in real-time using a partial least squares regression algorithm. Due to the inherent molecular selectivity for crystalline as well as organic matrix components, infrared spectroscopy uniquelyThe exposure of mining workers to crystalline particles, e.g., alpha quartz in respirable dust, is a ubiquitous global problem in occupational safety and health at surface and underground operations. The challenge of rapid in-field monitoring for direct assessment and adoption of intervention has not been solved satisfactorily to date, as conventional analytical methods such as X-ray diffraction and infrared spectroscopy require laboratory environments, complex system handling, tedious sample preparation, and are limited by, e.g., addressable particle size. A novel monitoring approach was developed for potential in-field application enabling the quantification of crystalline particles in the respirable regime based on transmission infrared spectroscopy. This on-site approach analyzes samples of dust in ambient air collected onto PVC filters using respirable dust sampling devices. In the present study, we demonstrate that portable Fourier transform infrared (FT-IR) spectroscopy in combination with multivariate data analysis provides a versatile tool for the identification and quantification of minerals in complex real-world matrices. Without further sample preparation, the loaded filters are immediately analyzed via transmission infrared spectroscopy, and the mineral amount is quantified in real-time using a partial least squares regression algorithm. Due to the inherent molecular selectivity for crystalline as well as organic matrix components, infrared spectroscopy uniquely allows to precisely determine the particle composition even in complex samples such as dust from coal mines or clay-rich environments. For establishing a robust partial least squares regression model, a method was developed for generating calibration samples representative in size and composition for respirable mine dust via aerodynamic size separation. Combined with experimental design strategies, this allows tailoring the calibration set to the demands of air quality management in underground mining scenarios, i.e., the respirable particle size regime and the matrix of the target analyte. … (more)
- Is Part Of:
- Applied spectroscopy. Volume 75:Issue 3(2021)
- Journal:
- Applied spectroscopy
- Issue:
- Volume 75:Issue 3(2021)
- Issue Display:
- Volume 75, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 75
- Issue:
- 3
- Issue Sort Value:
- 2021-0075-0003-0000
- Page Start:
- 307
- Page End:
- 316
- Publication Date:
- 2021-03
- Subjects:
- Fourier transform infrared spectroscopy -- FT-IR -- multivariate data analyses -- respirable particles -- mine dust -- occupational safety
Spectrum analysis -- Periodicals
543.505 - Journal URLs:
- http://asp.sagepub.com/ ↗
http://www.ingentaconnect.com/content/sas/sas ↗
http://www.uk.sagepub.com/home.nav ↗
http://firstsearch.oclc.org/journal=0003-7028;screen=info;ECOIP ↗ - DOI:
- 10.1177/0003702820968062 ↗
- Languages:
- English
- ISSNs:
- 0003-7028
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15194.xml