Dosimetric evaluation of gold nanoparticle aided intraoperative radiotherapy with the Intrabeam system using Monte Carlo simulations. (January 2021)
- Record Type:
- Journal Article
- Title:
- Dosimetric evaluation of gold nanoparticle aided intraoperative radiotherapy with the Intrabeam system using Monte Carlo simulations. (January 2021)
- Main Title:
- Dosimetric evaluation of gold nanoparticle aided intraoperative radiotherapy with the Intrabeam system using Monte Carlo simulations
- Authors:
- Moradi, F.
Abdul Sani, S.F.
Khandaker, M.U.
Sulieman, A.
Bradley, D.A. - Abstract:
- Abstract: Radiosensitization using high atomic number nanoparticles (NPs) has been shown to be an effective method to enhance radiotherapy efficiency. The pathways by which NPs cause sensitization, are generally categorized as physical, chemical and biological effects. Specifically in the case of keV photon radiotherapy where the contribution of physical effects in radiosensitization mechanism is considerable, Monte Carlo (MC) simulations have been an efficient tool to predict the radioenhancement level and to calculate dose enhancement factor (DEF). To-date, several analytical, simulational and experimental studies have reported the radiosensitization effect of gold nanoparticles (GNPs) in various brachytherapy situations. In this work we report for the first time, the DEFs achievable in intraoperative radiotherapy through use of the Intrabeam system and its spherical applicators with addition of GNPs. The MCNPX Monte Carlo code was used for radiation transport and dose calculations. The results of macroscopic and microscopic analysis show that for the Intrabeam system and a homogeneous distribution of 50 nm diameter GNPs, respective DEFs of up to some 1.5, 2, 2.5 and 3 in the tumour bed can be achieved with 5, 10, 15 and 20 mg/g concentrations. Due to rapid change in electron spectra, DEFs greater than 1 mm separation from the applicator surface decrease with distance, offering an additional advantage. Highlights: Gold nanoparticles (GNPs) are potential radiosensitizersAbstract: Radiosensitization using high atomic number nanoparticles (NPs) has been shown to be an effective method to enhance radiotherapy efficiency. The pathways by which NPs cause sensitization, are generally categorized as physical, chemical and biological effects. Specifically in the case of keV photon radiotherapy where the contribution of physical effects in radiosensitization mechanism is considerable, Monte Carlo (MC) simulations have been an efficient tool to predict the radioenhancement level and to calculate dose enhancement factor (DEF). To-date, several analytical, simulational and experimental studies have reported the radiosensitization effect of gold nanoparticles (GNPs) in various brachytherapy situations. In this work we report for the first time, the DEFs achievable in intraoperative radiotherapy through use of the Intrabeam system and its spherical applicators with addition of GNPs. The MCNPX Monte Carlo code was used for radiation transport and dose calculations. The results of macroscopic and microscopic analysis show that for the Intrabeam system and a homogeneous distribution of 50 nm diameter GNPs, respective DEFs of up to some 1.5, 2, 2.5 and 3 in the tumour bed can be achieved with 5, 10, 15 and 20 mg/g concentrations. Due to rapid change in electron spectra, DEFs greater than 1 mm separation from the applicator surface decrease with distance, offering an additional advantage. Highlights: Gold nanoparticles (GNPs) are potential radiosensitizers for use in brachytherapy. MC simulation is a familiar approach in prediction of dose enhancement factor (DEF). DEFs achievable using GNPs in Intrabeam system brachytherapy were estimated. Macroscopic and microscopic analysis were performed on the MC results. DEFs of 1.5, 2, 2.5 & 3 were obtained with 5, 10, 15 & 20 mg/g gold concentrations. … (more)
- Is Part Of:
- Radiation physics and chemistry. Volume 178(2021)
- Journal:
- Radiation physics and chemistry
- Issue:
- Volume 178(2021)
- Issue Display:
- Volume 178, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 178
- Issue:
- 2021
- Issue Sort Value:
- 2021-0178-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-01
- Subjects:
- Gold nanoparticle -- Monte Carlo simulation -- Dose enhancement -- Intraoperative radiotherapy -- Intrabeam
Radiation chemistry -- Periodicals
Radiometry -- Periodicals
Radiation -- Periodicals
Chimie sous rayonnement -- Périodiques
539.2 - Journal URLs:
- http://www.sciencedirect.com/science/journal/0969806X ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiation-physics-and-chemistry/ ↗ - DOI:
- 10.1016/j.radphyschem.2020.108864 ↗
- Languages:
- English
- ISSNs:
- 0969-806X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7227.984000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15172.xml