Evaluation of an a priori scatter correction algorithm for cone-beam computed tomography based range and dose calculations in proton therapy. (October 2020)
- Record Type:
- Journal Article
- Title:
- Evaluation of an a priori scatter correction algorithm for cone-beam computed tomography based range and dose calculations in proton therapy. (October 2020)
- Main Title:
- Evaluation of an a priori scatter correction algorithm for cone-beam computed tomography based range and dose calculations in proton therapy
- Authors:
- Andersen, Andreas Gravgaard
Park, Yang-Kyun
Elstrøm, Ulrik Vindelev
Petersen, Jørgen Breede Baltzer
Sharp, Gregory C.
Winey, Brian
Dong, Lei
Muren, Ludvig Paul - Abstract:
- Abstract: Background and purpose: Scatter correction of cone-beam computed tomography (CBCT) projections may enable accurate online dose-delivery estimations in photon and proton-based radiotherapy. This study aimed to evaluate the impact of scatter correction in CBCT-based proton range/dose calculations, in scans acquired in both proton and photon gantries. Material and methods: CBCT projections of a Catphan and an Alderson phantom were acquired on both a proton and a photon gantry. The scatter corrected CBCTs (corrCBCTs) and the clinical reconstructions (stdCBCTs) were compared against CTs rigidly registered to the CBCTs (rigidCTs). The CBCTs of the Catphan phantom were segmented by materials for CT number analysis. Water equivalent path length (WEPL) maps were calculated through the Alderson phantom while proton plans optimized on the rigidCT and recalculated on all CBCTs were compared in a gamma analysis. Results: In medium and high-density materials, the corrCBCT CT numbers were much closer to those of the rigidCT than the stdCBCTs. E.g. in the 50% bone segmentations the differences were reduced from above 300 HU (with stdCBCT) to around 60–70 HU (with corrCBCT). Differences in WEPL from the rigidCT were typically well below 5 mm for the corrCBCTs, compared to well above 10 mm for the stdCBCTs with the largest deviations in the head and thorax regions. Gamma pass rates (2%/2mm) when comparing CBCT-based dose re-calculations to rigidCT calculations were improved fromAbstract: Background and purpose: Scatter correction of cone-beam computed tomography (CBCT) projections may enable accurate online dose-delivery estimations in photon and proton-based radiotherapy. This study aimed to evaluate the impact of scatter correction in CBCT-based proton range/dose calculations, in scans acquired in both proton and photon gantries. Material and methods: CBCT projections of a Catphan and an Alderson phantom were acquired on both a proton and a photon gantry. The scatter corrected CBCTs (corrCBCTs) and the clinical reconstructions (stdCBCTs) were compared against CTs rigidly registered to the CBCTs (rigidCTs). The CBCTs of the Catphan phantom were segmented by materials for CT number analysis. Water equivalent path length (WEPL) maps were calculated through the Alderson phantom while proton plans optimized on the rigidCT and recalculated on all CBCTs were compared in a gamma analysis. Results: In medium and high-density materials, the corrCBCT CT numbers were much closer to those of the rigidCT than the stdCBCTs. E.g. in the 50% bone segmentations the differences were reduced from above 300 HU (with stdCBCT) to around 60–70 HU (with corrCBCT). Differences in WEPL from the rigidCT were typically well below 5 mm for the corrCBCTs, compared to well above 10 mm for the stdCBCTs with the largest deviations in the head and thorax regions. Gamma pass rates (2%/2mm) when comparing CBCT-based dose re-calculations to rigidCT calculations were improved from around 80% (with stdCBCT) to mostly above 90% (with corrCBCT). Conclusion: Scatter correction leads to substantial artefact reductions, improving accuracy of CBCT-based proton range/dose calculations. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 16(2020)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 16(2020)
- Issue Display:
- Volume 16, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 16
- Issue:
- 2020
- Issue Sort Value:
- 2020-0016-2020-0000
- Page Start:
- 89
- Page End:
- 94
- Publication Date:
- 2020-10
- Subjects:
- Cone beam -- CB -- Cone beam computed tomography -- CBCT -- Scatter -- Scatter correction -- Shading correction -- A priori -- Proton -- Range -- Dose calculation -- Dose recalculation -- Adaptive proton therapy -- APT -- Inter-fractional motion management -- Cone beam projections -- Projections -- Water equivalent path length -- WEPL -- Beam hardening
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2020.09.014 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15165.xml