Comparative Accuracy Analysis of a Real-time and an Intermittent-Scanning Continuous Glucose Monitoring System. (March 2021)
- Record Type:
- Journal Article
- Title:
- Comparative Accuracy Analysis of a Real-time and an Intermittent-Scanning Continuous Glucose Monitoring System. (March 2021)
- Main Title:
- Comparative Accuracy Analysis of a Real-time and an Intermittent-Scanning Continuous Glucose Monitoring System
- Authors:
- Link, Manuela
Kamecke, Ulrike
Waldenmaier, Delia
Pleus, Stefan
Garcia, Arturo
Haug, Cornelia
Freckmann, Guido - Abstract:
- Background: Currently, two different types of continuous glucose monitoring (CGM) systems are available: real time (rt) CGM systems that continuously provide glucose values and intermittent-scanning (is) CGM systems. This study compared accuracy of an rtCGM and an isCGM system when worn in parallel. Methods: Dexcom G5 Mobile (DG5) and FreeStyle Libre (FL) were worn in parallel by 27 subjects for 14 days including two clinic sessions with induced glucose excursions. The percentage of CGM values within ±20% or ±20 mg/dL of the laboratory comparison method results (YSI 2300 STAT Plus, YSI Inc., Yellow Springs, OH, United States; glucose oxidase based) or blood glucose meter values and mean absolute relative difference (MARD) were calculated. Consensus error grid and continuous glucose error grid analyses were performed to assess clinical accuracy. Results: Both systems displayed clinically accurate readings. Compared to laboratory comparison method results during clinic sessions, DG5 had 91.5% of values within ±20%/20 mg/dL and a MARD of 9.5%; FL had 82.5% of scanned values within ±20%/20 mg/dL and an MARD of 13.6%. Both systems showed a lower level of performance during the home phase and when using the blood glucose meter as reference. Conclusion: The two systems tested in this study represent two different principles of CGM. DG5 generally provided higher accordance with laboratory comparison method results than FL.
- Is Part Of:
- Journal of diabetes science and technology. Volume 15:Number 2(2021)
- Journal:
- Journal of diabetes science and technology
- Issue:
- Volume 15:Number 2(2021)
- Issue Display:
- Volume 15, Issue 2 (2021)
- Year:
- 2021
- Volume:
- 15
- Issue:
- 2
- Issue Sort Value:
- 2021-0015-0002-0000
- Page Start:
- 287
- Page End:
- 293
- Publication Date:
- 2021-03
- Subjects:
- accuracy -- continuous glucose monitoring -- mean absolute relative difference -- performance -- sensor
Diabetes -- Periodicals
Medical technology -- Periodicals
Diabetes Mellitus -- Periodicals
616.462005 - Journal URLs:
- http://ejournals.ebsco.com/direct.asp?JournalID=712321 ↗
http://www.jodsat.org/about.html ↗
http://online.sagepub.com/ ↗ - DOI:
- 10.1177/1932296819895022 ↗
- Languages:
- English
- ISSNs:
- 1932-2968
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15114.xml