The MB score: a new risk stratification index to predict the need for advanced tools in lead extraction procedures. Issue 4 (22nd February 2020)
- Record Type:
- Journal Article
- Title:
- The MB score: a new risk stratification index to predict the need for advanced tools in lead extraction procedures. Issue 4 (22nd February 2020)
- Main Title:
- The MB score: a new risk stratification index to predict the need for advanced tools in lead extraction procedures
- Authors:
- Bontempi, Luca
Curnis, Antonio
Della Bella, Paolo
Cerini, Manuel
Radinovic, Andrea
Inama, Lorenza
Melillo, Francesco
Salghetti, Francesca
Marzi, Alessandra
Gargaro, Alessio
Giacopelli, Daniele
Mazzone, Patrizio - Abstract:
- Abstract: Aims: A validated risk stratification schema for transvenous lead extraction (TLE) could improve the management of these procedures. We aimed to derive and validate a scoring system to efficiently predict the need for advanced tools to achieve TLE success. Methods and results: Between November 2013 and March 2018, 1960 leads were extracted in 973 consecutive TLE procedures in two national referral sites using a stepwise approach. A procedure was defined as advanced extraction if required the use of powered sheaths and/or snares. The study population was a posteriori 1:1 randomized in derivation and validation cohorts. In the derivation cohort, presence of more than two targeted leads (odds ratio [OR] 1.76, P = 0.049), 3-year-old (OR 3.04, P = 0.001), 5-year-old (OR 3.48, P < 0.001), 10-year-old (OR 3.58, P = 0.008) oldest lead, implantable cardioverter-defibrillator (OR 3.84, P < 0.001), and passive fixation lead (OR 1.91, P = 0.032) were selected by a stepwise procedure and constituted the MB score showing a C-statistics of 0.82. In the validation group, the MB score was significantly associated with the risk of advanced extraction (OR 2.40, 95% confidence interval 2.02-2.86, P < 0.001) and showed an increase in event rate with increasing score. A low value (threshold = 1) ensured 100% sensibility and 100% negative predictive value, while a high value (threshold = 5) allowed a specificity of 92.8% and a positive predictive value of 91.9%. Conclusion: InAbstract: Aims: A validated risk stratification schema for transvenous lead extraction (TLE) could improve the management of these procedures. We aimed to derive and validate a scoring system to efficiently predict the need for advanced tools to achieve TLE success. Methods and results: Between November 2013 and March 2018, 1960 leads were extracted in 973 consecutive TLE procedures in two national referral sites using a stepwise approach. A procedure was defined as advanced extraction if required the use of powered sheaths and/or snares. The study population was a posteriori 1:1 randomized in derivation and validation cohorts. In the derivation cohort, presence of more than two targeted leads (odds ratio [OR] 1.76, P = 0.049), 3-year-old (OR 3.04, P = 0.001), 5-year-old (OR 3.48, P < 0.001), 10-year-old (OR 3.58, P = 0.008) oldest lead, implantable cardioverter-defibrillator (OR 3.84, P < 0.001), and passive fixation lead (OR 1.91, P = 0.032) were selected by a stepwise procedure and constituted the MB score showing a C-statistics of 0.82. In the validation group, the MB score was significantly associated with the risk of advanced extraction (OR 2.40, 95% confidence interval 2.02-2.86, P < 0.001) and showed an increase in event rate with increasing score. A low value (threshold = 1) ensured 100% sensibility and 100% negative predictive value, while a high value (threshold = 5) allowed a specificity of 92.8% and a positive predictive value of 91.9%. Conclusion: In this study, we developed and tested a simple point-based scoring system able to efficiently identify patients at low and high risk of needing advanced tools during TLE procedures. … (more)
- Is Part Of:
- Europace. Volume 22:Issue 4(2020)
- Journal:
- Europace
- Issue:
- Volume 22:Issue 4(2020)
- Issue Display:
- Volume 22, Issue 4 (2020)
- Year:
- 2020
- Volume:
- 22
- Issue:
- 4
- Issue Sort Value:
- 2020-0022-0004-0000
- Page Start:
- 613
- Page End:
- 621
- Publication Date:
- 2020-02-22
- Subjects:
- Cardiac implantable electronic devices -- Lead extraction -- Lead extraction tools -- Implantable cardioverter-defibrillator -- Pacemaker -- MB score
Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euaa027 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15092.xml