Comparison of Icare Rebound Tonometer and Perkins Applanation Tonometer in Community Eye Screening. Issue 3 (May 2019)
- Record Type:
- Journal Article
- Title:
- Comparison of Icare Rebound Tonometer and Perkins Applanation Tonometer in Community Eye Screening. Issue 3 (May 2019)
- Main Title:
- Comparison of Icare Rebound Tonometer and Perkins Applanation Tonometer in Community Eye Screening
- Authors:
- Ting, Siew Leng
Lim, Lik Thai
Ooi, Chor Yau
Rahman, Md Mizanur - Abstract:
- Abstract : Purpose: To compare the measurement of intraocular pressure (IOP) of Icare rebound tonometer and Perkins applanation tonometer (PAT) during community eye screening and to assess the agreement between these 2 instruments. Design: A cross-sectional, non-interventional study. Methods: The IOP measurements by handheld Icare rebound tonometer (Finland) were first performed by a primary care physician. Then the IOP was measured using Perkins Mk3 applanation tonometer (Haag-Streit, UK) by an ophthalmologist who was masked to previous readings from the Icare rebound tonometer. The mean IOP measured by each tonometer was compared. Pearson correlation coefficient was used to explore the correlation between the IOP measurements of the 2 instruments. The level of agreement between them was assessed using the Bland and Altman method. Results: A total of 420 left eyes were examined. The mean age of subjects was 38.6 ± 18.2 years. Approximately 67% of subjects were female. The mean IOP was 16.3 ± 4.0 mm Hg using Icare and 13.4 ± 2.3 mm Hg using PAT. Pearson correlation coefficient showed a moderate positive correlation between the 2 methods ( r = +0.524, P < 0.001). Linear regression analysis revealed a slope of 0.28 with R2 of 0.255. The mean difference between the 2 methods was 2.90 ± 3.5 mm Hg and the sample t -test revealed a statistically significant mean difference from 0 ( P < 0.001). The 95% limits of agreement between the 2 methods were between -9.73 and 3.93 mm Hg.Abstract : Purpose: To compare the measurement of intraocular pressure (IOP) of Icare rebound tonometer and Perkins applanation tonometer (PAT) during community eye screening and to assess the agreement between these 2 instruments. Design: A cross-sectional, non-interventional study. Methods: The IOP measurements by handheld Icare rebound tonometer (Finland) were first performed by a primary care physician. Then the IOP was measured using Perkins Mk3 applanation tonometer (Haag-Streit, UK) by an ophthalmologist who was masked to previous readings from the Icare rebound tonometer. The mean IOP measured by each tonometer was compared. Pearson correlation coefficient was used to explore the correlation between the IOP measurements of the 2 instruments. The level of agreement between them was assessed using the Bland and Altman method. Results: A total of 420 left eyes were examined. The mean age of subjects was 38.6 ± 18.2 years. Approximately 67% of subjects were female. The mean IOP was 16.3 ± 4.0 mm Hg using Icare and 13.4 ± 2.3 mm Hg using PAT. Pearson correlation coefficient showed a moderate positive correlation between the 2 methods ( r = +0.524, P < 0.001). Linear regression analysis revealed a slope of 0.28 with R2 of 0.255. The mean difference between the 2 methods was 2.90 ± 3.5 mm Hg and the sample t -test revealed a statistically significant mean difference from 0 ( P < 0.001). The 95% limits of agreement between the 2 methods were between -9.73 and 3.93 mm Hg. Conclusions: The handheld Icare rebound tonometer is a reasonably acceptable screening tool in community practices. However, Icare overestimated IOP with a mean of 2.90 mm Hg higher than the PAT. Thus, using Goldmann applanation tonometer as a confirmatory measurement tool of IOP is suggested. … (more)
- Is Part Of:
- Asia-Pacific journal of ophthalmology. Volume 8:Issue 3(2019)
- Journal:
- Asia-Pacific journal of ophthalmology
- Issue:
- Volume 8:Issue 3(2019)
- Issue Display:
- Volume 8, Issue 3 (2019)
- Year:
- 2019
- Volume:
- 8
- Issue:
- 3
- Issue Sort Value:
- 2019-0008-0003-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-05
- Subjects:
- Icare rebound -- intraocular pressure -- Perkins tonometer
Ophthalmology -- Periodicals
Eye -- Diseases -- Periodicals
Periodicals
617.7005 - Journal URLs:
- http://journals.lww.com/apjoo/pages/default.aspx ↗
http://ovidsp.tx.ovid.com/sp-3.15.1b/ovidweb.cgi?S=ODEGFPELAADDOHBGNCKKOHFBBKLOAA00&TOC=S.sh.22.23.28.29&journal_browse_filter=jp|318 ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.22608/APO.2018433 ↗
- Languages:
- English
- ISSNs:
- 0129-1653
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 15032.xml