Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: A Retrieval Analysis. Issue 3 (October 2017)
- Record Type:
- Journal Article
- Title:
- Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: A Retrieval Analysis. Issue 3 (October 2017)
- Main Title:
- Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: A Retrieval Analysis
- Authors:
- Koch, Chelsea N.
Figgie, Mark
Figgie, Mark P.
Elpers, Marcella E.
Wright, Timothy M.
Padgett, Douglas E. - Abstract:
- Background: The BIOLOX® option system, consisting of a BIOLOX® delta ceramic femoral head with a titanium alloy adapter sleeve, is being increasingly utilized in revision hip arthroplasty. The sleeve protects the ceramic head from fracture and improper motion about the stem trunnion when a damaged trunnion is encountered at revision surgery. Corrosion and fretting due to metal-metal contact at the taper region of hip prosthesis create the potential of causing periprosthetic osteolysis and adverse local tissue reactions. Questions/Purposes: The objective of this study was to identify the type and extent of damage to retrieved sleeves and ceramic heads to determine their in vivo performance. Methods: Twenty-four ceramic heads with titanium alloy sleeves were examined. The articular and taper surfaces for each ceramic head were assessed for metal transfer using a subjective grading system. All surfaces of the 24 titanium sleeves and stem trunnions (only available for 7 of 24 cases) were assessed for corrosion and fretting using an established grading system. Scanning electron microscopy and energy dispersive X-ray analysis were conducted on representative sample of sleeves. Results: Fretting and corrosion were higher at the inner surface of the taper sleeve than the outer sleeve. Mean fretting scores at the inner taper and outer taper sleeve surfaces were 1.8 and 1.2, respectively. The mean corrosion score at the inner taper surface was 1.8; no corrosion was observed on theBackground: The BIOLOX® option system, consisting of a BIOLOX® delta ceramic femoral head with a titanium alloy adapter sleeve, is being increasingly utilized in revision hip arthroplasty. The sleeve protects the ceramic head from fracture and improper motion about the stem trunnion when a damaged trunnion is encountered at revision surgery. Corrosion and fretting due to metal-metal contact at the taper region of hip prosthesis create the potential of causing periprosthetic osteolysis and adverse local tissue reactions. Questions/Purposes: The objective of this study was to identify the type and extent of damage to retrieved sleeves and ceramic heads to determine their in vivo performance. Methods: Twenty-four ceramic heads with titanium alloy sleeves were examined. The articular and taper surfaces for each ceramic head were assessed for metal transfer using a subjective grading system. All surfaces of the 24 titanium sleeves and stem trunnions (only available for 7 of 24 cases) were assessed for corrosion and fretting using an established grading system. Scanning electron microscopy and energy dispersive X-ray analysis were conducted on representative sample of sleeves. Results: Fretting and corrosion were higher at the inner surface of the taper sleeve than the outer sleeve. Mean fretting scores at the inner taper and outer taper sleeve surfaces were 1.8 and 1.2, respectively. The mean corrosion score at the inner taper surface was 1.8; no corrosion was observed on the outer surface of any taper sleeve. SEM and EDS analyses provided further indications of low levels of damage. Conclusion: Fretting and corrosion were less severe than previously reported for conventional THA metal-metal taper connections, indicating that a ceramic head and titanium sleeve is a safe alternative in revision THA. … (more)
- Is Part Of:
- HSS journal. Volume 13:Issue 3(2017)
- Journal:
- HSS journal
- Issue:
- Volume 13:Issue 3(2017)
- Issue Display:
- Volume 13, Issue 3 (2017)
- Year:
- 2017
- Volume:
- 13
- Issue:
- 3
- Issue Sort Value:
- 2017-0013-0003-0000
- Page Start:
- 241
- Page End:
- 247
- Publication Date:
- 2017-10
- Subjects:
- total hip arthroplasty -- retrieval analysis -- fretting -- corrosion -- adapter sleeve
Musculoskeletal system -- Surgery -- Periodicals
Orthopedic surgery -- Periodicals
Musculoskeletal System -- surgery -- Periodicals
Orthopedic Procedures -- Periodicals
Orthopédie -- Périodiques
Appareil locomoteur -- Maladies -- Périodiques
Appareil locomoteur -- Maladies -- Patients -- Réadaptation -- Périodiques
617.47005 - Journal URLs:
- http://www.ncbi.nlm.nih.gov/pmc/journals/593 ↗
http://www.springerlink.com/content/1556-3316/ ↗
http://www.springerlink.com/openurl.asp?genre=journal&issn=1556-3316 ↗
http://www.springer.com/gb/ ↗ - DOI:
- 10.1007/s11420-017-9566-4 ↗
- Languages:
- English
- ISSNs:
- 1556-3316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4335.344650
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15029.xml