Humeral Tray-Taper Failure in Modular Reverse Total Shoulder Arthroplasty. Issue 1 (February 2016)
- Record Type:
- Journal Article
- Title:
- Humeral Tray-Taper Failure in Modular Reverse Total Shoulder Arthroplasty. Issue 1 (February 2016)
- Main Title:
- Humeral Tray-Taper Failure in Modular Reverse Total Shoulder Arthroplasty
- Authors:
- McDonald, Lucas S.
Dines, Joshua S.
Chin, Christopher
Warren, Russell F.
Dines, David M. - Abstract:
- Background: Reverse total shoulder arthroplasty (RTSA) provides reconstructive options in patients previously not candidates for total shoulder arthroplasty (TSA) or who have failed previous anatomic TSA. Revision from anatomic TSA to RTSA previously required removal of all components, a difficult and extensive procedure. Modular humeral components permit conversion from anatomic TSA to RTSA without removal of well-fixed humeral components. Questions/Purposes: Our purpose is to present a case series of patients treated for the unique and not previously reported complication of humeral tray-taper failure following modular RTSA. Challenges in diagnosis and treatment are described, including the use of dynamic fluoroscopy and manufacturer-specific instruments for component revision. Methods: Five patients with a total of six humeral tray-taper failures were identified from 300 patients with first-generation (titanium) humeral trays over a 7-year period. Dynamic fluoroscopic evaluation aided in diagnosis in a majority of the cases. All cases have been revised to second-generation (cobalt chrome) humeral trays. Results: Average follow-up was 22 ± 23 months (range 3–60 months). One individual required a second revision for the same complication, but otherwise, no additional procedures were required. Symptom relief was obtained in all patients. Conclusions: This case series illustrates a previously unpublished complication of humeral tray-taper junction failure following modularBackground: Reverse total shoulder arthroplasty (RTSA) provides reconstructive options in patients previously not candidates for total shoulder arthroplasty (TSA) or who have failed previous anatomic TSA. Revision from anatomic TSA to RTSA previously required removal of all components, a difficult and extensive procedure. Modular humeral components permit conversion from anatomic TSA to RTSA without removal of well-fixed humeral components. Questions/Purposes: Our purpose is to present a case series of patients treated for the unique and not previously reported complication of humeral tray-taper failure following modular RTSA. Challenges in diagnosis and treatment are described, including the use of dynamic fluoroscopy and manufacturer-specific instruments for component revision. Methods: Five patients with a total of six humeral tray-taper failures were identified from 300 patients with first-generation (titanium) humeral trays over a 7-year period. Dynamic fluoroscopic evaluation aided in diagnosis in a majority of the cases. All cases have been revised to second-generation (cobalt chrome) humeral trays. Results: Average follow-up was 22 ± 23 months (range 3–60 months). One individual required a second revision for the same complication, but otherwise, no additional procedures were required. Symptom relief was obtained in all patients. Conclusions: This case series illustrates a previously unpublished complication of humeral tray-taper junction failure following modular RTSA. Clinical and radiographic diagnosis is challenging; however, dynamic fluoroscopic evaluation permits identification of the component failure, and revision surgery results in good outcomes. We must, however, continue to evaluate what activities are recommend for patients following shoulder arthroplasty, specifically reverse total shoulder arthroplasty. … (more)
- Is Part Of:
- HSS journal. Volume 12:Issue 1(2016)
- Journal:
- HSS journal
- Issue:
- Volume 12:Issue 1(2016)
- Issue Display:
- Volume 12, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 12
- Issue:
- 1
- Issue Sort Value:
- 2016-0012-0001-0000
- Page Start:
- 8
- Page End:
- 12
- Publication Date:
- 2016-02
- Subjects:
- reverse total shoulder arthroplasty -- complication -- humeral tray-taper failure -- modular components -- revision surgery
Musculoskeletal system -- Surgery -- Periodicals
Orthopedic surgery -- Periodicals
Musculoskeletal System -- surgery -- Periodicals
Orthopedic Procedures -- Periodicals
Orthopédie -- Périodiques
Appareil locomoteur -- Maladies -- Périodiques
Appareil locomoteur -- Maladies -- Patients -- Réadaptation -- Périodiques
617.47005 - Journal URLs:
- http://www.ncbi.nlm.nih.gov/pmc/journals/593 ↗
http://www.springerlink.com/content/1556-3316/ ↗
http://www.springerlink.com/openurl.asp?genre=journal&issn=1556-3316 ↗
http://www.springer.com/gb/ ↗ - DOI:
- 10.1007/s11420-015-9470-8 ↗
- Languages:
- English
- ISSNs:
- 1556-3316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4335.344650
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15021.xml