Estimation of technical treatment accuracy in fractionated stereotactic radiosurgery. Issue 1 (29th April 2019)
- Record Type:
- Journal Article
- Title:
- Estimation of technical treatment accuracy in fractionated stereotactic radiosurgery. Issue 1 (29th April 2019)
- Main Title:
- Estimation of technical treatment accuracy in fractionated stereotactic radiosurgery
- Authors:
- Bisht, Raj Kishor
Natanasabapathi, Gopishankar
Sharad Kale, Shashank - Abstract:
- Abstract: Aim: The purpose of this study was to estimate technical treatment accuracy in fractionated stereotactic radiosurgery (fSRS) using the Extend™ system (ES) of Gamma Knife (GK). Methods and materials: The fSRS with GK relies on a re-locatable ES where the reference treatment position is estimated using repositioning check tool (RCT). A patient surveillance unit (PSU) monitors the head and neck movement of the patient during treatment and imaging. The quality assurance test of RCT was performed to evaluate a standard error (SE) associated with a measurement tool called digital probe. A '4-mm collimator shot' dose plan for a head–neck phantom was investigated using EBT3 films. CT and MR distortion measurement studies were combined to evaluate SEimaging . The combined uncertainty from all measurements was evaluated using statistical methods, and the resultant treatment accuracy was investigated for the ES. Results: Four sets of RCT measurements and 20 observations of associated digital probe showed SERCT of ±0·0186 mm and SEdigital probe of ±0·0002 mm. The mean positional shift of 0·2752 mm ( σ = 0·0696 mm) was observed for 20 treatment settings of the phantom. The differences between radiological and predefined isocentres were 0·4650 and 0·4270 mm for two independent experiments. SEimaging and SEdiode tool were evaluated as ±0·1055 and ±0·0096 mm, respectively. An expanded uncertainty of ±0·2371 mm (at 95% confidence level) was observed with our system. Conclusions:Abstract: Aim: The purpose of this study was to estimate technical treatment accuracy in fractionated stereotactic radiosurgery (fSRS) using the Extend™ system (ES) of Gamma Knife (GK). Methods and materials: The fSRS with GK relies on a re-locatable ES where the reference treatment position is estimated using repositioning check tool (RCT). A patient surveillance unit (PSU) monitors the head and neck movement of the patient during treatment and imaging. The quality assurance test of RCT was performed to evaluate a standard error (SE) associated with a measurement tool called digital probe. A '4-mm collimator shot' dose plan for a head–neck phantom was investigated using EBT3 films. CT and MR distortion measurement studies were combined to evaluate SEimaging . The combined uncertainty from all measurements was evaluated using statistical methods, and the resultant treatment accuracy was investigated for the ES. Results: Four sets of RCT measurements and 20 observations of associated digital probe showed SERCT of ±0·0186 mm and SEdigital probe of ±0·0002 mm. The mean positional shift of 0·2752 mm ( σ = 0·0696 mm) was observed for 20 treatment settings of the phantom. The differences between radiological and predefined isocentres were 0·4650 and 0·4270 mm for two independent experiments. SEimaging and SEdiode tool were evaluated as ±0·1055 and ±0·0096 mm, respectively. An expanded uncertainty of ±0·2371 mm (at 95% confidence level) was observed with our system. Conclusions: The combined result of the positional shift and expanded uncertainty showed close agreement with film investigations. … (more)
- Is Part Of:
- Journal of radiotherapy in practice. Volume 19:Issue 1(2020)
- Journal:
- Journal of radiotherapy in practice
- Issue:
- Volume 19:Issue 1(2020)
- Issue Display:
- Volume 19, Issue 1 (2020)
- Year:
- 2020
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2020-0019-0001-0000
- Page Start:
- 30
- Page End:
- 37
- Publication Date:
- 2019-04-29
- Subjects:
- expanded uncertainty, -- fractionated stereotactic radiosurgery, -- Gamma Knife, -- phantom
Radiotherapy -- Periodicals
615.842005 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=JRP ↗
- DOI:
- 10.1017/S1460396919000128 ↗
- Languages:
- English
- ISSNs:
- 1460-3969
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 14985.xml