The design and physical characterization of a multileaf collimator for robotic radiosurgery. (10th February 2016)
- Record Type:
- Journal Article
- Title:
- The design and physical characterization of a multileaf collimator for robotic radiosurgery. (10th February 2016)
- Main Title:
- The design and physical characterization of a multileaf collimator for robotic radiosurgery
- Authors:
- Asmerom, G
Bourne, D
Chappelow, J
Goggin, L M
Heitz, R
Jordan, P
Kilby, W
Laing, T
Maurer, C R
Noll, J M
Sayeh, S
Weber, A - Abstract:
- Abstract: A multileaf collimator (MLC) optimized for SBRT delivery with the CyberKnife ® Robotic Radiosurgery System (Accuray Incorporated, Sunnyvale, CA, USA) is described. The MLC is exchangeable with the alternate fixed and variable circular aperture collimator systems. The non-coplanar workspace is effectively equivalent for all three collimation types. The same range of tracking options, including real-time respiratory motion tracking, and the same tolerance on beam pointing accuracy (≤0.95 mm) is maintained with all three collimation types. The MLC includes 52 flat-sided leaves, each of which is 90 mm tall and projects 3.85 mm width at the nominal treatment distance of 800 mm SAD. The design allows 100% overtravel and unrestricted interdigitation. Leaf position is determined by primary motor encoders and is checked with a secondary optical camera system. Maximum leakage, including inter-leaf and under the closed position leaf-tip gap was measured on five units to be ≤0.44%, while mean leakage and transmission ranged from 0.22%–0.25%. Leaf positioning accuracy measured over the full range of leaf positions, all robot and MLC orientations, and including variation with leaf motion direction and accumulated leaf motion after initialization had a mean error <0.2 mm, with 2%–98% range of ≤±0.5 mm (projected at 800 mm SAD) on three units tested. The only factor found to effect leaf positioning accuracy was sag under gravity, which systematically altered leaf positions by 0.1Abstract: A multileaf collimator (MLC) optimized for SBRT delivery with the CyberKnife ® Robotic Radiosurgery System (Accuray Incorporated, Sunnyvale, CA, USA) is described. The MLC is exchangeable with the alternate fixed and variable circular aperture collimator systems. The non-coplanar workspace is effectively equivalent for all three collimation types. The same range of tracking options, including real-time respiratory motion tracking, and the same tolerance on beam pointing accuracy (≤0.95 mm) is maintained with all three collimation types. The MLC includes 52 flat-sided leaves, each of which is 90 mm tall and projects 3.85 mm width at the nominal treatment distance of 800 mm SAD. The design allows 100% overtravel and unrestricted interdigitation. Leaf position is determined by primary motor encoders and is checked with a secondary optical camera system. Maximum leakage, including inter-leaf and under the closed position leaf-tip gap was measured on five units to be ≤0.44%, while mean leakage and transmission ranged from 0.22%–0.25%. Leaf positioning accuracy measured over the full range of leaf positions, all robot and MLC orientations, and including variation with leaf motion direction and accumulated leaf motion after initialization had a mean error <0.2 mm, with 2%–98% range of ≤±0.5 mm (projected at 800 mm SAD) on three units tested. The only factor found to effect leaf positioning accuracy was sag under gravity, which systematically altered leaf positions by 0.1 mm. Tilting the leaves to reduce inter-leaf leakage results in 0.5 mm asymmetry in leaf-side penumbra at 100 mm depth, and a partial leaf-edge transmission pattern analogous to the tongue and groove effect observed with interlocking leaves. … (more)
- Is Part Of:
- Biomedical physics & engineering express. Volume 2:Number 1(2016)
- Journal:
- Biomedical physics & engineering express
- Issue:
- Volume 2:Number 1(2016)
- Issue Display:
- Volume 2, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 2
- Issue:
- 1
- Issue Sort Value:
- 2016-0002-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-02-10
- Subjects:
- x-ray beam collimation -- CyberKnife® System -- robotic radiosurgery -- stereotactic body radiation therapy -- multileaf collimator
Medical physics -- Periodicals
Biophysics -- Periodicals
Biomedical engineering -- Periodicals
Medical sciences -- Periodicals
610.153 - Journal URLs:
- http://iopscience.iop.org/2057-1976/ ↗
http://www.iop.org/ ↗ - DOI:
- 10.1088/2057-1976/2/1/017003 ↗
- Languages:
- English
- ISSNs:
- 2057-1976
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 14978.xml