Automated multi-parameter high-dose-rate brachytherapy quality assurance via radioluminescence imaging. (11th November 2020)
- Record Type:
- Journal Article
- Title:
- Automated multi-parameter high-dose-rate brachytherapy quality assurance via radioluminescence imaging. (11th November 2020)
- Main Title:
- Automated multi-parameter high-dose-rate brachytherapy quality assurance via radioluminescence imaging
- Authors:
- Jia, Mengyu
Kim, Tae Jin
Yang, Yong
Xing, Lei
Jean, Paul De
Grafil, Elliot
Jenkins, Cesare H
Fahimian, Benjamin P - Abstract:
- Abstract: The purpose of this study is to leverage radioluminescence imaging for the development of an automated high-dose-rate (HDR) brachytherapy quality assurance (QA) system that enables simultaneous measurements of dwell position, dwell time, wire velocity, and relative source strength in a single test. The system consists of a radioluminescence phosphor sheet (a mixture of Gd2 O2 S:Tb and PDMS) positioned atop a HDR needle applicator, a complementary metal-oxide-semiconductor digital camera used to capture the emitted radioluminescence signals from the scintillator sheet, and an in-house graphical user interface for signal processing. The signal processing was used to extract source intensity, location, and elapsed time, yielding the final measurements on dwell position, dwell time, and wire velocity. The source strength relative to the well chamber calibration (in unit of Air-Kerma strength, Sk ) is measured by establishing a calibration curve that correlates Sk with the detector response. Validation experiments are performed using three customized treatment plans. With these plans, the dwell position and dwell time are verified for a range of 110.0 cm–117.5 cm and 2 s–16 s, respectively, and the linear correlation with Sk is demonstrated for the source strength varying between 28 348 U (cGy cm 2 h −1 ) and 41 906 U. The wire velocity, i.e. the speed of the radioactive source averaged over the distance in between dwell positions, is calculated for various distancesAbstract: The purpose of this study is to leverage radioluminescence imaging for the development of an automated high-dose-rate (HDR) brachytherapy quality assurance (QA) system that enables simultaneous measurements of dwell position, dwell time, wire velocity, and relative source strength in a single test. The system consists of a radioluminescence phosphor sheet (a mixture of Gd2 O2 S:Tb and PDMS) positioned atop a HDR needle applicator, a complementary metal-oxide-semiconductor digital camera used to capture the emitted radioluminescence signals from the scintillator sheet, and an in-house graphical user interface for signal processing. The signal processing was used to extract source intensity, location, and elapsed time, yielding the final measurements on dwell position, dwell time, and wire velocity. The source strength relative to the well chamber calibration (in unit of Air-Kerma strength, Sk ) is measured by establishing a calibration curve that correlates Sk with the detector response. Validation experiments are performed using three customized treatment plans. With these plans, the dwell position and dwell time are verified for a range of 110.0 cm–117.5 cm and 2 s–16 s, respectively, and the linear correlation with Sk is demonstrated for the source strength varying between 28 348 U (cGy cm 2 h −1 ) and 41 906 U. The wire velocity, i.e. the speed of the radioactive source averaged over the distance in between dwell positions, is calculated for various distances ranging from 5 mm to 50 mm. Results show that the mean deviations of the measured dwell position and dwell time are 0.1 mm (range from 0 to 0.2 mm) and 32.5 ms (range from 0 to 60.0 ms) with respect to the planned values, respectively, and the system response is highly linear with Sk ( R 2 = 0.998). Moreover, the measured wire velocities are comparable to previously reported values. Benefitting from the compact hardware design and image processing algorithms, the system provides a practical, reliable, and comprehensive solution for HDR QA. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 65:Number 22(2020:Nov.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 65:Number 22(2020:Nov.)
- Issue Display:
- Volume 65, Issue 22 (2020)
- Year:
- 2020
- Volume:
- 65
- Issue:
- 22
- Issue Sort Value:
- 2020-0065-0022-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-11-11
- Subjects:
- radioluminescence imaging -- quality assurance (QA) -- high dose rate (HDR) brachytherapy
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/abb570 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 14969.xml