Assessment of a standalone photoplethysmography (PPG) algorithm for detection of atrial fibrillation on wristband-derived data. (December 2020)
- Record Type:
- Journal Article
- Title:
- Assessment of a standalone photoplethysmography (PPG) algorithm for detection of atrial fibrillation on wristband-derived data. (December 2020)
- Main Title:
- Assessment of a standalone photoplethysmography (PPG) algorithm for detection of atrial fibrillation on wristband-derived data
- Authors:
- Selder, JL
Proesmans, T
Breukel, L
Dur, O
Gielen, W
van Rossum, AC
Allaart, CP - Abstract:
- Highlights: The fibricheck AF algorithm can accurately detect AF from wristband-derived PPG data in an elderly population, thus showing feasibility of non-invasive long-term AF screening and monitoring. Accuracy of AF detection by this PPG algorithm/wristband combination is comparable with 1-lead ECG devices. To prevent false positives, PPG algorithms need to be very strict in quality assessment. In this study 24% of the recordings are marked as insufficient quality. Abstract: Introduction: Atrial fibrillation (AF) is the most common cardiac arrhythmia in the developed world. Using photoplethysmography (PPG) and software algorithms, AF can be detected with high accuracy using smartphone camera-derived data. However, reports of diagnostic accuracy of standalone algorithms using wristband-derived PPG data are sparse, while this provides a means to perform long-term AF screening and monitoring. This study evaluated the diagnostic accuracy of a well-known standalone algorithm using wristband-derived PPG data. Materials and Methods: Subjects recruited from a community senior care organization were instructed to wear the Wavelet PPG wristband on one arm and the Alivecor KardiaBand one-lead-ECG wristband on the other. Three consecutive measurements (duration per measurement: 60 s for PPG and 30 s for one-lead ECG) were performed with both devices, simultaneously. The PPG data were analyzed by the Fibricheck standalone algorithm and the ECG data by the Kardia algorithm. The resultsHighlights: The fibricheck AF algorithm can accurately detect AF from wristband-derived PPG data in an elderly population, thus showing feasibility of non-invasive long-term AF screening and monitoring. Accuracy of AF detection by this PPG algorithm/wristband combination is comparable with 1-lead ECG devices. To prevent false positives, PPG algorithms need to be very strict in quality assessment. In this study 24% of the recordings are marked as insufficient quality. Abstract: Introduction: Atrial fibrillation (AF) is the most common cardiac arrhythmia in the developed world. Using photoplethysmography (PPG) and software algorithms, AF can be detected with high accuracy using smartphone camera-derived data. However, reports of diagnostic accuracy of standalone algorithms using wristband-derived PPG data are sparse, while this provides a means to perform long-term AF screening and monitoring. This study evaluated the diagnostic accuracy of a well-known standalone algorithm using wristband-derived PPG data. Materials and Methods: Subjects recruited from a community senior care organization were instructed to wear the Wavelet PPG wristband on one arm and the Alivecor KardiaBand one-lead-ECG wristband on the other. Three consecutive measurements (duration per measurement: 60 s for PPG and 30 s for one-lead ECG) were performed with both devices, simultaneously. The PPG data were analyzed by the Fibricheck standalone algorithm and the ECG data by the Kardia algorithm. The results were compared to a reference standard (interpretation of the one-lead ECG by two independent cardiologists). Results: A total of 180 PPGs and one-lead ECGs were recorded in 60 subjects, with a mean age of 70±17. AF was identified in 6 (10%) of the users, two users (3%) were not classifiable by the PPG algorithm and 1 user (2%) was not classifiable by the one-lead ECG algorithm. The diagnostic performance (sensitivity/specificity/positive predictive value/negative predictive value/accuracy) on user level was 100/96/75/100/97% for the PPG wristband and 100/98/86/100/98% for the one-lead ECG wristband. Conclusions: In a small real-world cohort of elderly people, the standalone Fibricheck AF algorithm can accurately detect AF using Wavelet wristband-derived PPG data. Results are comparable to the Alivecor Kardia one-lead ECG device, with an acceptable unclassifiable/bad quality rate. This opens the door for long-term AF screening and monitoring. … (more)
- Is Part Of:
- Computer methods and programs in biomedicine. Volume 197(2020)
- Journal:
- Computer methods and programs in biomedicine
- Issue:
- Volume 197(2020)
- Issue Display:
- Volume 197, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 197
- Issue:
- 2020
- Issue Sort Value:
- 2020-0197-2020-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-12
- Subjects:
- Smartwatch -- Wristband -- Atrial fibrillation -- Photoplethysmography -- One-lead ECG -- Algorithm -- Screening -- Monitoring -- Long-term -- Motion artefacts -- Quality assessment
Medicine -- Computer programs -- Periodicals
Biology -- Computer programs -- Periodicals
Computers -- Periodicals
Medicine -- Periodicals
Médecine -- Logiciels -- Périodiques
Biologie -- Logiciels -- Périodiques
Biology -- Computer programs
Medicine -- Computer programs
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/01692607 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.cmpb.2020.105753 ↗
- Languages:
- English
- ISSNs:
- 0169-2607
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3394.095000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14946.xml