Experimental Analysis of the Quality of Needle-Assisted Fenestration in Aortic Stent-Grafts and the Differences Between Gradual and Rapid Balloon Dilation. (February 2021)
- Record Type:
- Journal Article
- Title:
- Experimental Analysis of the Quality of Needle-Assisted Fenestration in Aortic Stent-Grafts and the Differences Between Gradual and Rapid Balloon Dilation. (February 2021)
- Main Title:
- Experimental Analysis of the Quality of Needle-Assisted Fenestration in Aortic Stent-Grafts and the Differences Between Gradual and Rapid Balloon Dilation
- Authors:
- Li, Dong-lin
Zeng, Qing-long
Xiang, Yi-lang
Qiu, Chen-yang
Li, Zhen-jiang
He, Yang-yan
Zhu, Qian-qian
Wu, Zi-heng
Wang, Xun
Zhang, Hong-kun - Abstract:
- Purpose: To report the findings of an in vitro experiment to evaluate the quality of needle fenestrations dilated by different size balloons in various stent-grafts and to investigate the differences between gradual and rapid dilation.Materials and Methods: Fenestrations were made using an 18-G needle in 5 different polyester or expanded polytetrafluoroethylene (ePTFE) stent-grafts: Relay, Valiant, Hercules, TAG, and Ankura. Each stent-graft received 2 groups of fenestrations: one was followed by gradual sequential dilation (4-, 6-, 8-, and 10-mm balloons) and the other by rapid dilation (4- and 10-mm balloons). The pressure was increased to 10 atmospheres or until the balloon was fully inflated with no waist. Quantitative and qualitative evaluations, including fenestration diameter, area, shape, and margins were conducted using light microscopy and scanning electron microscopy.Results: Relay had the strongest resistance to dilation and Ankura the slightest. The maximum length and area of holes expanded as the balloon diameter increased. The fenestrations in polyester devices were mostly elliptical or slit-like, with limited tears but extensive fibers visible in the margin, while ePTFE stent-grafts showed larger fenestration areas with clearer margins. Ankura showed the best quality of fenestrations, which were always circular or square without fabric tears, while the holes in the TAG were square or elliptical but sometimes had a slit after large balloon dilation (≥6 mm).Purpose: To report the findings of an in vitro experiment to evaluate the quality of needle fenestrations dilated by different size balloons in various stent-grafts and to investigate the differences between gradual and rapid dilation.Materials and Methods: Fenestrations were made using an 18-G needle in 5 different polyester or expanded polytetrafluoroethylene (ePTFE) stent-grafts: Relay, Valiant, Hercules, TAG, and Ankura. Each stent-graft received 2 groups of fenestrations: one was followed by gradual sequential dilation (4-, 6-, 8-, and 10-mm balloons) and the other by rapid dilation (4- and 10-mm balloons). The pressure was increased to 10 atmospheres or until the balloon was fully inflated with no waist. Quantitative and qualitative evaluations, including fenestration diameter, area, shape, and margins were conducted using light microscopy and scanning electron microscopy.Results: Relay had the strongest resistance to dilation and Ankura the slightest. The maximum length and area of holes expanded as the balloon diameter increased. The fenestrations in polyester devices were mostly elliptical or slit-like, with limited tears but extensive fibers visible in the margin, while ePTFE stent-grafts showed larger fenestration areas with clearer margins. Ankura showed the best quality of fenestrations, which were always circular or square without fabric tears, while the holes in the TAG were square or elliptical but sometimes had a slit after large balloon dilation (≥6 mm). The Relay, Valiant, Hercules, and Ankura devices showed no difference in maximum diameter, fenestration area, or scores of shape and margin (p>0.05). Rapid dilation in the TAG increased the rate of uncontrolled fabric tear, resulting in a larger final diameter (12.90 vs 10.82 mm, p=0.047), smaller area (30.46 vs 41.09 mm 2, p=0.028), worse shape (0.75 vs 1.20, p=0.268), and worse margin (0.40 vs 1.00, p=0.174). Though the decreased fenestration shape and margin scores did not reach statistical significance, the trend for decline was more obvious than with the other devices.Conclusion: Materials and structures of the stent-grafts determine the quality of fenestrations dilated by different size balloons. The use of sequential vs rapid balloon dilation is also crucial for fashioning high-quality fenestrations and should be selected judiciously. … (more)
- Is Part Of:
- Journal of endovascular therapy. Volume 28:Number 1(2021)
- Journal:
- Journal of endovascular therapy
- Issue:
- Volume 28:Number 1(2021)
- Issue Display:
- Volume 28, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 28
- Issue:
- 1
- Issue Sort Value:
- 2021-0028-0001-0000
- Page Start:
- 44
- Page End:
- 52
- Publication Date:
- 2021-02
- Subjects:
- aortic arch -- balloon dilation -- expanded polytetrafluoroethylene fabric -- fenestration quality -- in situ fenestration -- in vitro experiment -- needle fenestration -- polyester fabric -- stent-graft
Blood-vessels -- Endoscopic surgery -- Periodicals
Angioscopy -- Periodicals
Intravenous catheterization -- Periodicals
Peripheral vascular diseases -- Treatment -- Periodicals
Vascular Surgical Procedures -- Periodicals
Angioscopy -- Periodicals
Catheterization, Peripheral -- Periodicals
Peripheral Vascular Diseases -- therapy -- Periodicals
Angioscopie
Maladies vasculaires périphériques
617.413 - Journal URLs:
- http://jet.sagepub.com/ ↗
http://www.jevt.org ↗
http://www.uk.sagepub.com ↗ - DOI:
- 10.1177/1526602820947095 ↗
- Languages:
- English
- ISSNs:
- 1526-6028
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14845.xml