Can different Catphan phantoms be used in a multi-centre audit of radiotherapy CT image quality?. (October 2020)
- Record Type:
- Journal Article
- Title:
- Can different Catphan phantoms be used in a multi-centre audit of radiotherapy CT image quality?. (October 2020)
- Main Title:
- Can different Catphan phantoms be used in a multi-centre audit of radiotherapy CT image quality?
- Authors:
- Davis, Anne T.
Palmer, Antony L.
Nisbet, Andrew - Abstract:
- Highlights: Catphan phantoms of the same model have good consistency. Inter-phantom variability is much better than specification range suggests. Catphan 604 model have higher measured contrast than 504 model phantoms. Abstract: Purpose: To determine the variation between Catphan image quality CT phantoms, specifically for use in a future multi-centre image quality audit. Method: 14 Catphan phantoms (models 503, 504 and 604) were scanned on a Canon Aquilion Prime CT scanner using a single scan protocol. Measurements were made of noise in the uniformity section, visibility of low contrast targets and contrast, x-ray attenuation and CT number for 5 materials in the sensitometry section. Scans were also acquired using one phantom and varying reconstruction field of view, image slice thickness, effective tube-current-time product and iterative reconstruction settings to determine how the degree of inter-phantom variability compared with the magnitude of changes from scan parameter alteration. Results: Across all phantoms the mean CT value in the uniformity section was 7.0 (SD 0.9) range: 4.9–8.1 HU. For the different materials the CT numbers were air: −1004 ± 5, Polymethylpentene: −190 ± 2, Polystyrene: −42 ± 2, Delrin: 321 ± 5 and Teflon: 898 ± 8 HU. Consistency of low contrast targets through visual scoring was good. Measured contrast was lower ( p < 0.001) with more variability for 504 versus 604 models. All phantoms produced identical tube current settings with x-ray tubeHighlights: Catphan phantoms of the same model have good consistency. Inter-phantom variability is much better than specification range suggests. Catphan 604 model have higher measured contrast than 504 model phantoms. Abstract: Purpose: To determine the variation between Catphan image quality CT phantoms, specifically for use in a future multi-centre image quality audit. Method: 14 Catphan phantoms (models 503, 504 and 604) were scanned on a Canon Aquilion Prime CT scanner using a single scan protocol. Measurements were made of noise in the uniformity section, visibility of low contrast targets and contrast, x-ray attenuation and CT number for 5 materials in the sensitometry section. Scans were also acquired using one phantom and varying reconstruction field of view, image slice thickness, effective tube-current-time product and iterative reconstruction settings to determine how the degree of inter-phantom variability compared with the magnitude of changes from scan parameter alteration. Results: Across all phantoms the mean CT value in the uniformity section was 7.0 (SD 0.9) range: 4.9–8.1 HU. For the different materials the CT numbers were air: −1004 ± 5, Polymethylpentene: −190 ± 2, Polystyrene: −42 ± 2, Delrin: 321 ± 5 and Teflon: 898 ± 8 HU. Consistency of low contrast targets through visual scoring was good. Measured contrast was lower ( p < 0.001) with more variability for 504 versus 604 models. All phantoms produced identical tube current settings with x-ray tube current modulation, indicating no x-ray attenuation differences. The degree of change in image quality metrics between phantoms was small compared with results when scan parameters were varied. Conclusion: Catphan phantoms model 604 showed minimal differences and will be used for multi-centre inter-comparison work, with the consistency between phantoms appropriate for measuring possible variations in image quality. … (more)
- Is Part Of:
- Physica medica. Volume 78(2020)
- Journal:
- Physica medica
- Issue:
- Volume 78(2020)
- Issue Display:
- Volume 78, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 78
- Issue:
- 2020
- Issue Sort Value:
- 2020-0078-2020-0000
- Page Start:
- 38
- Page End:
- 47
- Publication Date:
- 2020-10
- Subjects:
- Catphan -- Phantom -- Consistency -- CT
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2020.09.003 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14812.xml