Accuracy and Precision of an Ultrasound‐Based Device to Measure the Pelvic Tilt in Several Positions. (30th October 2019)
- Record Type:
- Journal Article
- Title:
- Accuracy and Precision of an Ultrasound‐Based Device to Measure the Pelvic Tilt in Several Positions. (30th October 2019)
- Main Title:
- Accuracy and Precision of an Ultrasound‐Based Device to Measure the Pelvic Tilt in Several Positions
- Authors:
- Dardenne, Guillaume
Pluchon, Jean‐Philippe
Letissier, Hoel
Guezou‐Philippe, Aziliz
Gerard, Romain
Lefèvre, Christian
Stindel, Eric - Abstract:
- Abstract : Objectives: A new ultrasound‐based device is proposed to measure the patient‐specific pelvic tilt in different daily positions. The aim of this study was to assess the accuracy of this device as well as the intraobserver and interobserver precisions. Methods: The accuracy was assessed by performing several tilt measurements with the device on a testing mechanical bench. The error was defined as the difference between the tilt measured with the device and the tilt provided by this test bench. Three physicians, a novice, an intermediate, and an expert user, were also asked to perform 10 measurements on 3 healthy volunteers with low, medium, and high body mass indices to analyze the intraobserver and interobserver precisions. These 10 measurements were performed in the standing, sitting, and supine positions. Results: The mean accuracy of the device ± SD was 1.1° ± 0.7° (range, 0°–4.0°). The interobserver and intraobserver precisions were excellent whatever the body mass index and good to excellent according to the positions. There was no learning curve, and the time required to complete the measurements was approximately 5 minutes. Conclusions: This study presents an accurate and precise noninvasive device for measurement of the pelvic tilt in different positions.
- Is Part Of:
- Journal of ultrasound in medicine. Volume 39:Number 4(2020)
- Journal:
- Journal of ultrasound in medicine
- Issue:
- Volume 39:Number 4(2020)
- Issue Display:
- Volume 39, Issue 4 (2020)
- Year:
- 2020
- Volume:
- 39
- Issue:
- 4
- Issue Sort Value:
- 2020-0039-0004-0000
- Page Start:
- 667
- Page End:
- 674
- Publication Date:
- 2019-10-30
- Subjects:
- pelvic tilt -- total hip arthroplasty -- ultrasound‐based measurement device
Ultrasonics in medicine -- Periodicals
Ultrasonics
Ultrasonography
Ultrasonics in medicine
Electronic journals
Periodicals
Periodicals
616.07543 - Journal URLs:
- http://www.jultrasoundmed.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jum.15141 ↗
- Languages:
- English
- ISSNs:
- 0278-4297
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5071.455000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: