Analytical validation of eight methods of thyroglobulin measurement in fine-needle aspiration washouts. (January 2021)
- Record Type:
- Journal Article
- Title:
- Analytical validation of eight methods of thyroglobulin measurement in fine-needle aspiration washouts. (January 2021)
- Main Title:
- Analytical validation of eight methods of thyroglobulin measurement in fine-needle aspiration washouts
- Authors:
- Boux de Casson, Florence
Beloeil, Rémi
Gauchez, Anne-Sophie
Oris, Charlotte
Leban, Monique
Schlageter, Marie-Hélène
Moineau, Marie-Pierre
Dufour-Rainfray, Diane
Bach-Ngohou, Kalyane
Chikh, Karim
Moal, Valérie - Abstract:
- Background: Thyroglobulin (Tg) assay in washout fluids of fine needles, after cervical lymph nodes aspiration, is used for detecting metastases from differentiated thyroid carcinomas. Assay methods are the same as for Tg in serum. However, with non-serum samples, methods require extensive validation to notably check for the absence of matrix effect. This study fits this context. Our objectives were to assess analytic performances, in washout fluid, of eight different Tg assay methods and to compare them to validated data in serum. Methods: Eleven medical laboratories participated in this study. The matrix tested was phosphate-buffer saline containing 1% bovine serum albumin (PBS-1% BSA). Samples used were dilutions, in this buffer, of Certified Reference Material (CRM 457). We verified, for all methods, the limit of detection, precision, linearity, trueness and accuracy. Results: In PBS-1% BSA, the functional sensitivities (FS) were comparable to those expected for serum. All the methods were linear. The relative biases of trueness were between –24.5 and 10.2% around 1 µ g/L. Total analytical error was ≤40% near the functional sensitivity values. Conclusion: No quantitatively important matrix effect was observed. All the methods showed their ability to measure Tg in PBS-1% BSA, over the concentration range of interest, with acceptable total analytical error. We validated the functional sensitivity value as a decision threshold in thyroidectomized patients after treatmentBackground: Thyroglobulin (Tg) assay in washout fluids of fine needles, after cervical lymph nodes aspiration, is used for detecting metastases from differentiated thyroid carcinomas. Assay methods are the same as for Tg in serum. However, with non-serum samples, methods require extensive validation to notably check for the absence of matrix effect. This study fits this context. Our objectives were to assess analytic performances, in washout fluid, of eight different Tg assay methods and to compare them to validated data in serum. Methods: Eleven medical laboratories participated in this study. The matrix tested was phosphate-buffer saline containing 1% bovine serum albumin (PBS-1% BSA). Samples used were dilutions, in this buffer, of Certified Reference Material (CRM 457). We verified, for all methods, the limit of detection, precision, linearity, trueness and accuracy. Results: In PBS-1% BSA, the functional sensitivities (FS) were comparable to those expected for serum. All the methods were linear. The relative biases of trueness were between –24.5 and 10.2% around 1 µ g/L. Total analytical error was ≤40% near the functional sensitivity values. Conclusion: No quantitatively important matrix effect was observed. All the methods showed their ability to measure Tg in PBS-1% BSA, over the concentration range of interest, with acceptable total analytical error. We validated the functional sensitivity value as a decision threshold in thyroidectomized patients after treatment and with low concentrations of serum Tg. … (more)
- Is Part Of:
- Annals of clinical biochemistry. Volume 58:Number 1(2021)
- Journal:
- Annals of clinical biochemistry
- Issue:
- Volume 58:Number 1(2021)
- Issue Display:
- Volume 58, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 58
- Issue:
- 1
- Issue Sort Value:
- 2021-0058-0001-0000
- Page Start:
- 54
- Page End:
- 65
- Publication Date:
- 2021-01
- Subjects:
- Thyroglobulin -- immunoassay -- thyroid disease
Clinical chemistry -- Periodicals
Clinical biochemistry -- Periodicals
616.075 - Journal URLs:
- http://web.ebscohost.com/ehost/detail?sid=810a7788-77dd-439f-9630-ad7f5b199fd3%40sessionmgr4&vid=1&hid=14&bdata=JnNpdGU9ZWhvc3QtbGl2ZSZzY29wZT1zaXRl#db=mnh&jid=0324055 ↗
http://acb.rsmjournals.com ↗
http://www.usc.edu/hsc/nml/e-resources/info/annclib.html ↗
http://www.uk.sagepub.com/home.nav ↗
http://www.ingentaconnect.com/content/rsm/acb ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1177/0004563220968369 ↗
- Languages:
- English
- ISSNs:
- 0004-5632
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 14733.xml