The optimal incubation time for in vitro hemocompatibility testing: Assessment using polymer reference materials under pulsatile flow with physiological wall shear stress conditions. Issue 7 (29th January 2019)
- Record Type:
- Journal Article
- Title:
- The optimal incubation time for in vitro hemocompatibility testing: Assessment using polymer reference materials under pulsatile flow with physiological wall shear stress conditions. Issue 7 (29th January 2019)
- Main Title:
- The optimal incubation time for in vitro hemocompatibility testing: Assessment using polymer reference materials under pulsatile flow with physiological wall shear stress conditions
- Authors:
- Blok, Sjoerd Leendert Johannes
van Oeveren, Willem
Engels, Gerwin Erik - Abstract:
- Abstract: During hemocompatibility testing, activation products may reach plateau values which can result in less distinction between hemocompatible and hemo‐incompatible materials. Of concern is an underestimation of the blood activation caused by the biomaterial of interest, which may result in a false assessment of hemocompatibility. To elucidate the optimal incubation time for in vitro hemocompatibility testing, we used the Haemobile circulation model with human whole blood. Blood from healthy volunteers was in vitro incubated under pulsatile flow with physiological wall shear stress conditions at 37°C for 30, 60, 120, or 240 min. Test loops containing low‐density polyethylene and polydimethylsiloxane served as low and high reference materials, that is, hemocompatible and hemo‐incompatible biomaterials, respectively. In addition, empty loops served as a negative reference. Thrombogenicity, platelet function, inflammatory response, coagulation, and hemolysis between references and incubation times were compared. We found that thrombogenicity and platelet function were significantly affected by both the duration of incubation and the type of material. In particular, thrombogenicity and platelet function assessments were affected by incubation time. We found that an exposure time of 60 min was sufficient, and for almost all variables an optimal incubation time to discriminate between the low and high reference material. © 2019 The Authors. Journal of Biomedical MaterialsAbstract: During hemocompatibility testing, activation products may reach plateau values which can result in less distinction between hemocompatible and hemo‐incompatible materials. Of concern is an underestimation of the blood activation caused by the biomaterial of interest, which may result in a false assessment of hemocompatibility. To elucidate the optimal incubation time for in vitro hemocompatibility testing, we used the Haemobile circulation model with human whole blood. Blood from healthy volunteers was in vitro incubated under pulsatile flow with physiological wall shear stress conditions at 37°C for 30, 60, 120, or 240 min. Test loops containing low‐density polyethylene and polydimethylsiloxane served as low and high reference materials, that is, hemocompatible and hemo‐incompatible biomaterials, respectively. In addition, empty loops served as a negative reference. Thrombogenicity, platelet function, inflammatory response, coagulation, and hemolysis between references and incubation times were compared. We found that thrombogenicity and platelet function were significantly affected by both the duration of incubation and the type of material. In particular, thrombogenicity and platelet function assessments were affected by incubation time. We found that an exposure time of 60 min was sufficient, and for almost all variables an optimal incubation time to discriminate between the low and high reference material. © 2019 The Authors. Journal of Biomedical Materials Research Part B: Applied Biomaterials published by Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 107B: 2335–2342, 2019. … (more)
- Is Part Of:
- Journal of biomedical materials research. Volume 107:Issue 7(2019)
- Journal:
- Journal of biomedical materials research
- Issue:
- Volume 107:Issue 7(2019)
- Issue Display:
- Volume 107, Issue 7 (2019)
- Year:
- 2019
- Volume:
- 107
- Issue:
- 7
- Issue Sort Value:
- 2019-0107-0007-0000
- Page Start:
- 2335
- Page End:
- 2342
- Publication Date:
- 2019-01-29
- Subjects:
- biomaterials -- blood–material interaction -- cardiovascular device -- hemocompatibility -- hemocompatibility testing
Biomedical materials -- Periodicals
610.28 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jbm.b.34326 ↗
- Languages:
- English
- ISSNs:
- 1552-4973
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.725000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14575.xml