Prediction of human observer performance in a 2‐alternative forced choice low‐contrast detection task using channelized Hotelling observer: Impact of radiation dose and reconstruction algorithms. Issue 4 (18th March 2013)
- Record Type:
- Journal Article
- Title:
- Prediction of human observer performance in a 2‐alternative forced choice low‐contrast detection task using channelized Hotelling observer: Impact of radiation dose and reconstruction algorithms. Issue 4 (18th March 2013)
- Main Title:
- Prediction of human observer performance in a 2‐alternative forced choice low‐contrast detection task using channelized Hotelling observer: Impact of radiation dose and reconstruction algorithms
- Authors:
- Yu, Lifeng
Leng, Shuai
Chen, Lingyun
Kofler, James M.
Carter, Rickey E.
McCollough, Cynthia H. - Abstract:
- Abstract : Purpose: : Efficient optimization of CT protocols demands a quantitative approach to predicting human observer performance on specific tasks at various scan and reconstruction settings. The goal of this work was to investigate how well a channelized Hotelling observer (CHO) can predict human observer performance on 2‐alternative forced choice (2AFC) lesion‐detection tasks at various dose levels and two different reconstruction algorithms: a filtered‐backprojection (FBP) and an iterative reconstruction (IR) method. Methods: : A 35 × 26 cm 2 torso‐shaped phantom filled with water was used to simulate an average‐sized patient. Three rods with different diameters (small: 3 mm; medium: 5 mm; large: 9 mm) were placed in the center region of the phantom to simulate small, medium, and large lesions. The contrast relative to background was −15 HU at 120 kV. The phantom was scanned 100 times using automatic exposure control each at 60, 120, 240, 360, and 480 quality reference mAs on a 128‐slice scanner. After removing the three rods, the water phantom was again scanned 100 times to provide signal‐absent background images at the exact same locations. By extracting regions of interest around the three rods and on the signal‐absent images, the authors generated 21 2AFC studies. Each 2AFC study had 100 trials, with each trial consisting of a signal‐present image and a signal‐absent image side‐by‐side in randomized order. In total, 2100 trials were presented to both the modelAbstract : Purpose: : Efficient optimization of CT protocols demands a quantitative approach to predicting human observer performance on specific tasks at various scan and reconstruction settings. The goal of this work was to investigate how well a channelized Hotelling observer (CHO) can predict human observer performance on 2‐alternative forced choice (2AFC) lesion‐detection tasks at various dose levels and two different reconstruction algorithms: a filtered‐backprojection (FBP) and an iterative reconstruction (IR) method. Methods: : A 35 × 26 cm 2 torso‐shaped phantom filled with water was used to simulate an average‐sized patient. Three rods with different diameters (small: 3 mm; medium: 5 mm; large: 9 mm) were placed in the center region of the phantom to simulate small, medium, and large lesions. The contrast relative to background was −15 HU at 120 kV. The phantom was scanned 100 times using automatic exposure control each at 60, 120, 240, 360, and 480 quality reference mAs on a 128‐slice scanner. After removing the three rods, the water phantom was again scanned 100 times to provide signal‐absent background images at the exact same locations. By extracting regions of interest around the three rods and on the signal‐absent images, the authors generated 21 2AFC studies. Each 2AFC study had 100 trials, with each trial consisting of a signal‐present image and a signal‐absent image side‐by‐side in randomized order. In total, 2100 trials were presented to both the model and human observers. Four medical physicists acted as human observers. For the model observer, the authors used a CHO with Gabor channels, which involves six channel passbands, five orientations, and two phases, leading to a total of 60 channels. The performance predicted by the CHO was compared with that obtained by four medical physicists at each 2AFC study. Results: : The human and model observers were highly correlated at each dose level for each lesion size for both FBP and IR. The Pearson's product‐moment correlation coefficients were 0.986 [95% confidence interval (CI): 0.958–0.996] for FBP and 0.985 (95% CI: 0.863–0.998) for IR. Bland‐Altman plots showed excellent agreement for all dose levels and lesions sizes with a mean absolute difference of 1.0% ± 1.1% for FBP and 2.1% ± 3.3% for IR. Conclusions: : Human observer performance on a 2AFC lesion detection task in CT with a uniform background can be accurately predicted by a CHO model observer at different radiation dose levels and for both FBP and IR methods. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 4(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 4(2013)
- Issue Display:
- Volume 40, Issue 4 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 4
- Issue Sort Value:
- 2013-0040-0004-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-03-18
- Subjects:
- Computed tomography -- Numerical optimization -- Dosimetry/exposure assessment -- Contrast -- Reconstruction
computerised tomography -- dosimetry -- filtering theory -- image reconstruction -- iterative methods -- medical image processing -- optimisation -- phantoms
computed tomography (CT) -- model observer -- image quality -- radiation dose -- iterative reconstruction (IR)
Computerised tomographs -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general
Medical imaging -- Computed tomography -- Medical image reconstruction -- Medical image quality -- Dosimetry -- Medical image noise -- Modulation transfer functions -- Nuclear medicine imaging -- Biomedical modeling -- Infrared radiation
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4794498 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14534.xml