Warming blood products for transfusion to neonates: In vitro assessments. Issue 9 (10th August 2020)
- Record Type:
- Journal Article
- Title:
- Warming blood products for transfusion to neonates: In vitro assessments. Issue 9 (10th August 2020)
- Main Title:
- Warming blood products for transfusion to neonates: In vitro assessments
- Authors:
- Hulse, Whitley
Bahr, Timothy M.
Fredrickson, Lauren
Canfield, Christy M.
Friddle, Kim
Pysher, Theodore J.
Ilstrup, Sarah J.
Ohls, Robin K.
Christensen, Robert D. - Abstract:
- Abstract: Background: Blood products may be transfused into neonates at temperatures at or below room temperature. The benefits and risks of warming blood to 37°C are not defined in this population or with the equipment used in neonates. Physiologic warming might enhance product effectiveness or decrease transfusion‐associated hypothermia. Study Design and Methods: We utilized an in vitro model of neonatal transfusions, with a syringe pump, blood tubing, and 24‐gauge catheter and compared current practice (cold products) vs an inline blood warmer. Transfusions were performed rapidly (30 minutes) and slower (120 minutes) to model emergent vs routine situations. We tested red blood cells, fresh‐frozen plasma, apheresis platelets (PLTs), and cold‐stored low‐titer group O whole blood. We used infrared detectors and inline probes to measure temperatures at the origin and at the simulated patient. We assessed warmer‐induced damage by measuring plasma hemoglobin and hematocrit (seeking hemolysis), fibrinogen (seeking activation of coagulation), and PLT count and TEG‐MA (seeking PLT destruction or dysfunction). Results: The cold‐stored products were 4.2 ± 1.0°C (mean ± SD) at the origin and 21.5 ± 0.1°C at the patient. With the inline warmer, products were 37.8 ± 0.6°C at the warmer and 32.6 ± 1.7°C at the patient during a 30‐minute infusion, but were 34.5 ± 2.1 with a foil sheath covering the terminal tubing. We found no warmer‐induced damage using any metric. Conclusion: InAbstract: Background: Blood products may be transfused into neonates at temperatures at or below room temperature. The benefits and risks of warming blood to 37°C are not defined in this population or with the equipment used in neonates. Physiologic warming might enhance product effectiveness or decrease transfusion‐associated hypothermia. Study Design and Methods: We utilized an in vitro model of neonatal transfusions, with a syringe pump, blood tubing, and 24‐gauge catheter and compared current practice (cold products) vs an inline blood warmer. Transfusions were performed rapidly (30 minutes) and slower (120 minutes) to model emergent vs routine situations. We tested red blood cells, fresh‐frozen plasma, apheresis platelets (PLTs), and cold‐stored low‐titer group O whole blood. We used infrared detectors and inline probes to measure temperatures at the origin and at the simulated patient. We assessed warmer‐induced damage by measuring plasma hemoglobin and hematocrit (seeking hemolysis), fibrinogen (seeking activation of coagulation), and PLT count and TEG‐MA (seeking PLT destruction or dysfunction). Results: The cold‐stored products were 4.2 ± 1.0°C (mean ± SD) at the origin and 21.5 ± 0.1°C at the patient. With the inline warmer, products were 37.8 ± 0.6°C at the warmer and 32.6 ± 1.7°C at the patient during a 30‐minute infusion, but were 34.5 ± 2.1 with a foil sheath covering the terminal tubing. We found no warmer‐induced damage using any metric. Conclusion: In simulated neonatal intensive care unit (NICU) transfusions, an inline blood warmer can deliver blood products at near‐physiologic temperatures with no detected damage. We suggest in vivo testing of warmed NICU transfusions, assessing product effectiveness and hypothermia risk reduction. … (more)
- Is Part Of:
- Transfusion. Volume 60:Issue 9(2020)
- Journal:
- Transfusion
- Issue:
- Volume 60:Issue 9(2020)
- Issue Display:
- Volume 60, Issue 9 (2020)
- Year:
- 2020
- Volume:
- 60
- Issue:
- 9
- Issue Sort Value:
- 2020-0060-0009-0000
- Page Start:
- 1924
- Page End:
- 1928
- Publication Date:
- 2020-08-10
- Subjects:
- blood warmer -- cold‐stored low titer group O whole blood -- fresh frozen plasma -- neonates -- platelets -- red blood cells -- safety
Hematology -- Periodicals
Blood -- Transfusion -- Periodicals
Blood Group Antigens -- Periodicals
Blood Preservation -- Periodicals
Blood Transfusion -- Periodicals
615 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1537-2995 ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=trf ↗
http://www.transfusion.org ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/trf.16007 ↗
- Languages:
- English
- ISSNs:
- 0041-1132
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9020.704000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 14356.xml