It is time to implement prepump arterial pressure monitoring during hemodialysis: A retrospective multicenter study. Issue 6 (November 2020)
- Record Type:
- Journal Article
- Title:
- It is time to implement prepump arterial pressure monitoring during hemodialysis: A retrospective multicenter study. Issue 6 (November 2020)
- Main Title:
- It is time to implement prepump arterial pressure monitoring during hemodialysis: A retrospective multicenter study
- Authors:
- Sun, Chun-Yan
Mo, Ya-Wen
Lan, Li-Juan
Han, Xiao-Wei
Song, Li
Zhang, Guan-Rong
Zhou, Li-Fang
Zheng, Shu-Qian
Chen, Ying-Gui
Liu, Shuang-Xin
Liang, Xin-Ling
Fu, Xia - Abstract:
- Introduction: Prepump arterial pressure (Pa) indicates the ease or difficulty with which the blood pump can draw blood from vascular access (inflow) during hemodialysis. The absolute prepump arterial pressure to blood pump speed (Qb) ratio (|Pa/Qb|) may reflect the dysfunction of other vascular accesses. There is no consensus on the impact of |Pa/Qb| on arteriovenous fistula dysfunction. This study aimed to demonstrate the impact of |Pa/Qb| on arteriovenous fistula dysfunction. Methods: In this retrospective analysis, 490 hemodialysis patients with arteriovenous fistula from three hospitals were enrolled. Data were extracted from the I-Diapro database and hospital case systems. The absolute values for |Pa/Qb| and other data collected in the first month of enrollment were used to predict arteriovenous fistula dysfunction and determine the |Pa/Qb| cutoff value. Based on this value, patients were grouped, and 1-year arteriovenous fistula function was analyzed. Patients were followed until arteriovenous fistula dysfunction, until access type replacement, or for 12 months. Results: The area under the receiver operating characteristic curve for fistula dysfunction over 1 year was 0.65, with an optimal |Pa/Qb| value, sensitivity, and specificity of 0.499, 60.7%, and 72.6%, respectively. |Pa/Qb| > 0.499 was associated with earlier intervention (317.37 ± 7.68 vs 345.96 ± 3.64 days), lower survival ( p < 0.001), and a 3.26-fold greater risk of arteriovenous fistula dysfunction ( pIntroduction: Prepump arterial pressure (Pa) indicates the ease or difficulty with which the blood pump can draw blood from vascular access (inflow) during hemodialysis. The absolute prepump arterial pressure to blood pump speed (Qb) ratio (|Pa/Qb|) may reflect the dysfunction of other vascular accesses. There is no consensus on the impact of |Pa/Qb| on arteriovenous fistula dysfunction. This study aimed to demonstrate the impact of |Pa/Qb| on arteriovenous fistula dysfunction. Methods: In this retrospective analysis, 490 hemodialysis patients with arteriovenous fistula from three hospitals were enrolled. Data were extracted from the I-Diapro database and hospital case systems. The absolute values for |Pa/Qb| and other data collected in the first month of enrollment were used to predict arteriovenous fistula dysfunction and determine the |Pa/Qb| cutoff value. Based on this value, patients were grouped, and 1-year arteriovenous fistula function was analyzed. Patients were followed until arteriovenous fistula dysfunction, until access type replacement, or for 12 months. Results: The area under the receiver operating characteristic curve for fistula dysfunction over 1 year was 0.65, with an optimal |Pa/Qb| value, sensitivity, and specificity of 0.499, 60.7%, and 72.6%, respectively. |Pa/Qb| > 0.499 was associated with earlier intervention (317.37 ± 7.68 vs 345.96 ± 3.64 days), lower survival ( p < 0.001), and a 3.26-fold greater risk of arteriovenous fistula dysfunction ( p < 0.001) than |Pa/Qb| ⩽ 0.499. Conclusions: |Pa/Qb| was an independent risk factor for arteriovenous fistula dysfunction. Nurses should emphasize |Pa/Qb| monitoring and properly set blood pump speed according to this ratio to protect arteriovenous fistula function. |Pa/Qb| > 0.499 might be a predictive measure of arteriovenous fistula dysfunction. … (more)
- Is Part Of:
- Journal of vascular access. Volume 21:Issue 6(2020)
- Journal:
- Journal of vascular access
- Issue:
- Volume 21:Issue 6(2020)
- Issue Display:
- Volume 21, Issue 6 (2020)
- Year:
- 2020
- Volume:
- 21
- Issue:
- 6
- Issue Sort Value:
- 2020-0021-0006-0000
- Page Start:
- 938
- Page End:
- 944
- Publication Date:
- 2020-11
- Subjects:
- Prepump arterial pressure -- blood flow rate -- pressure monitoring -- hemodialysis nursing
Arterial catheterization -- Periodicals
Intravenous catheterization -- Periodicals
612.13 - Journal URLs:
- http://journals.sagepub.com/home/jva ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/1129729820917266 ↗
- Languages:
- English
- ISSNs:
- 1129-7298
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14275.xml