A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy. (January 2018)
- Record Type:
- Journal Article
- Title:
- A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy. (January 2018)
- Main Title:
- A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy
- Authors:
- Lechner, Wolfgang
Wesolowska, Paulina
Azangwe, Godfrey
Arib, Mehenna
Alves, Victor Gabriel Leandro
Suming, Luo
Ekendahl, Daniela
Bulski, Wojciech
Samper, José Luis Alonso
Vinatha, Sumanth Panyam
Siri, Srimanoroth
Tomsej, Milan
Tenhunen, Mikko
Povall, Julie
Kry, Stephen F.
Followill, David S.
Thwaites, David I.
Georg, Dietmar
Izewska, Joanna - Abstract:
- Abstract: Background and purpose: An audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy was developed and tested through a multinational research group and performed on a national level in five different countries. Materials and methods: Centres participating in this study were asked to provide OFs calculated by their TPSs for 10 × 10 cm 2, 6 × 6 cm 2, 4 × 4 cm 2, 3 × 3 cm 2 and 2 × 2 cm 2 field sizes using an SSD of 100 cm. The ratio of these calculated OFs to reference OFs was analysed. The action limit was ±3% for the 2 × 2 cm 2 field and ±2% for all other fields. Results: OFs for more than 200 different beams were collected in total. On average, the OFs for small fields calculated by TPSs were generally larger than measured reference data. These deviations increased with decreasing field size. On a national level, 30% and 31% of the calculated OFs of the 2 × 2 cm 2 field exceeded the action limit of 3% for nominal beam energies of 6 MV and for nominal beam energies higher than 6 MV, respectively. Conclusion: Modern TPS beam models generally overestimate the OFs for small fields. The verification of calculated small field OFs is a vital step and should be included when commissioning a TPS. The methodology outlined in this study can be used to identify potential discrepancies in clinical beam models.
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 5(2018)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 5(2018)
- Issue Display:
- Volume 5, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 5
- Issue:
- 2018
- Issue Sort Value:
- 2018-0005-2018-0000
- Page Start:
- 58
- Page End:
- 63
- Publication Date:
- 2018-01
- Subjects:
- Dosimetry audit -- Small field output factors -- Treatment planning system
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2018.02.005 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14199.xml