Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy. (April 2018)
- Record Type:
- Journal Article
- Title:
- Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy. (April 2018)
- Main Title:
- Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
- Authors:
- Wyatt, Jonathan
Hedley, Stephen
Johnstone, Emily
Speight, Richard
Kelly, Charles
Henry, Ann
Short, Susan
Murray, Louise
Sebag-Montefiore, David
McCallum, Hazel - Abstract:
- Abstract: Background and purpose: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. Materials and methods: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. Results: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was < 0.4 mm for all scanners and sequences. For the 1 mm lateral offset scan 90 % of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%). Conclusions: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity toAbstract: Background and purpose: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. Materials and methods: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. Results: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was < 0.4 mm for all scanners and sequences. For the 1 mm lateral offset scan 90 % of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%). Conclusions: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 6(2018)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 6(2018)
- Issue Display:
- Volume 6, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 6
- Issue:
- 2018
- Issue Sort Value:
- 2018-0006-2018-0000
- Page Start:
- 31
- Page End:
- 38
- Publication Date:
- 2018-04
- Subjects:
- Geometric distortion -- Magnetic resonance imaging -- Radiotherapy -- MR-only radiotherapy planning
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2018.04.005 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14201.xml