Polarized multichannel transmit MRI to reduce shading near metal implants. Issue 1 (13th February 2015)
- Record Type:
- Journal Article
- Title:
- Polarized multichannel transmit MRI to reduce shading near metal implants. Issue 1 (13th February 2015)
- Main Title:
- Polarized multichannel transmit MRI to reduce shading near metal implants
- Authors:
- Bachschmidt, Theresa J.
Köhler, Michael
Nistler, Jürgen
Geppert, Christian
Jakob, Peter M.
Nittka, Mathias - Abstract:
- Abstract : Purpose: To investigate the benefit of a two‐channel transmit system on shading close to total hip replacements and other elongated metal structures in parallel to the magnet bore. Methods: An analytical model comprising a water cylinder and a metal rod is introduced to describe the B1 effects close to elongated metal structures and it is verified. The dependence of the optimal polarization, which induces minimum shading, on the position of the metal is analyzed. Furthermore, the optimal polarization for two patients is determined both on the basis of the model and experimentally and its benefit compared with circular polarization is investigated. Results: The cylindrical model approximates the modification of the B1 field due to the metal well, and the optimal polarization strongly depends on the position of the rod. In vivo, shading can be ameliorated by the use of the optimal polarization; for total hip replacements with shafts of titanium, this polarization can be determined on the basis of both the analytical model and the experimental data. Conclusion: Parallel transmission offers the possibility of a substantial reduction of shading close to long metal structures in parallel to the magnet bore. Magn Reson Med 75:217–226, 2016. © 2015 Wiley Periodicals, Inc.
- Is Part Of:
- Magnetic resonance in medicine. Volume 75:Issue 1(2016:Jan.)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 75:Issue 1(2016:Jan.)
- Issue Display:
- Volume 75, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 75
- Issue:
- 1
- Issue Sort Value:
- 2016-0075-0001-0000
- Page Start:
- 217
- Page End:
- 226
- Publication Date:
- 2015-02-13
- Subjects:
- hip replacement -- titanium -- shading -- parallel transmission -- 3T
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.25621 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14170.xml