Technical Note: Impact of region of interest size and location in Gafchromic film dosimetry. Issue 5 (18th April 2018)
- Record Type:
- Journal Article
- Title:
- Technical Note: Impact of region of interest size and location in Gafchromic film dosimetry. Issue 5 (18th April 2018)
- Main Title:
- Technical Note: Impact of region of interest size and location in Gafchromic film dosimetry
- Authors:
- Gholizadeh Sendani, Neda
Karimian, Alireza
Ferreira, Clara
Alaei, Parham - Abstract:
- Abstract : Purpose: To evaluate dependence of measured dose on size and location of region of interest (ROI) in Gafchromic EBT3 film dosimetry. Methods: Gafchromic EBT3 films were irradiated perpendicularly using the 6MV beam from a linear accelerator at 10 cm depth (100 cm SSD) of a 30 × 30 × 20 cm 3 solid water phantom for a range of field sizes of 6 × 6 to 100 × 100 mm 2 . ImageJ software was used for reading pieces of film. The appropriate location of ROIs in scanned films was found by two methods. First, the ROI was visually placed at the center of image. Second, the profile of pixel value versus distance was plotted and the center of profile was used for drawing ROI. Each scanned film was read using both methods and for three ROI sizes (1, 2, and 4 mm). A plastic scintillator, Exradin W1, was used as the reference dosimeter. Results: Comparing the three ROI sizes using both methods showed that there was less than 2% difference from reference in output factor measurements for field sizes larger or equal to 10 × 10 mm 2 . The percentage differences were increased in field sizes smaller than 10 × 10 mm 2 and for ROI size of 4 × 4 mm 2 for both centered‐ROI and profiled‐ROI methods. The mean percentage differences from reference measurements, for field sizes of 100 × 100 to 20 × 20 mm 2, were smaller than 1% in both methods of ROI positioning. For field sizes of 15 × 15 and 10 × 10 mm 2, the smaller mean percentage differences were observed in profiled‐ROI (4 × 4 mm 2 )Abstract : Purpose: To evaluate dependence of measured dose on size and location of region of interest (ROI) in Gafchromic EBT3 film dosimetry. Methods: Gafchromic EBT3 films were irradiated perpendicularly using the 6MV beam from a linear accelerator at 10 cm depth (100 cm SSD) of a 30 × 30 × 20 cm 3 solid water phantom for a range of field sizes of 6 × 6 to 100 × 100 mm 2 . ImageJ software was used for reading pieces of film. The appropriate location of ROIs in scanned films was found by two methods. First, the ROI was visually placed at the center of image. Second, the profile of pixel value versus distance was plotted and the center of profile was used for drawing ROI. Each scanned film was read using both methods and for three ROI sizes (1, 2, and 4 mm). A plastic scintillator, Exradin W1, was used as the reference dosimeter. Results: Comparing the three ROI sizes using both methods showed that there was less than 2% difference from reference in output factor measurements for field sizes larger or equal to 10 × 10 mm 2 . The percentage differences were increased in field sizes smaller than 10 × 10 mm 2 and for ROI size of 4 × 4 mm 2 for both centered‐ROI and profiled‐ROI methods. The mean percentage differences from reference measurements, for field sizes of 100 × 100 to 20 × 20 mm 2, were smaller than 1% in both methods of ROI positioning. For field sizes of 15 × 15 and 10 × 10 mm 2, the smaller mean percentage differences were observed in profiled‐ROI (4 × 4 mm 2 ) and centered‐ROI (4 × 4 mm 2 ). For the field sizes of 8 × 8 and 6 × 6 mm 2, the profiled‐ROI (2 × 2 mm 2 ) had smallest mean percentage difference, which was 0.88%. Conclusion: The ROI size of 4 × 4 mm 2 is appropriate for dose measurements in field sizes of 100 × 100 mm 2 to 10 × 10 mm 2, regardless of the method of finding location of ROI. In field sizes smaller than 10 × 10 mm 2, finding location of the ROI by profile of pixel values increases the accuracy of measurement, and ROI size of 2 × 2 mm 2 has the smallest difference from the reference dose measurements. … (more)
- Is Part Of:
- Medical physics. Volume 45:Issue 5(2018)
- Journal:
- Medical physics
- Issue:
- Volume 45:Issue 5(2018)
- Issue Display:
- Volume 45, Issue 5 (2018)
- Year:
- 2018
- Volume:
- 45
- Issue:
- 5
- Issue Sort Value:
- 2018-0045-0005-0000
- Page Start:
- 2329
- Page End:
- 2336
- Publication Date:
- 2018-04-18
- Subjects:
- film dosimetry -- Gafchromic film -- region of interest -- small field
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.12885 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14170.xml