Zilberti, Luca - Abstract:
- Abstract: This work focuses on the in silico evaluation of the energy deposed by MRI switched gradient fields in bulk metallic implants and the consequent temperature increase in the surrounding tissues. An original computational strategy, based on the subdivision of the gradient coil switching sequences into sub-signals and on the time-harmonic electromagnetic field solution, allows to realistically simulate the evolution of the phenomena produced by the gradient coils fed according to any MRI sequence. Then, Pennes' bioheat equation is solved through a Douglas–Gunn time split scheme to compute the time-dependent temperature increase. The procedure is validated by comparison with laboratory results, using a component of a realistic hip implant embedded within a phantom, obtaining an agreement on the temperature increase better than 5%, lower than the overall measurement uncertainty. The heating generated inside the body of a patient with a unilateral hip implant when undergoing an Echo-Planar Imaging (EPI) MRI sequence is evaluated and the role of the parameters affecting the thermal results (body position, coil performing the frequency encoding, effects of thermoregulation) is discussed. The results show that the gradient coils can generate local increases of temperature up to some kelvin when acting without radiofrequency excitation. Hence, their contribution in general should not be disregarded when evaluating patients' safety.
- Is Part Of:
- Physics in medicine & biology. Volume 64:Number 24(2019:Dec.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 64:Number 24(2019:Dec.)
- Issue Display:
- Volume 64, Issue 24 (2019)
- Year:
- 2019
- Volume:
- 64
- Issue:
- 24
- Issue Sort Value:
- 2019-0064-0024-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-12-13
- Subjects:
- dosimetry -- magnetic resonance imaging (MRI) -- MR safety -- gradient coils -- medical implants -- prostheses -- numerical simulation
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/ab5428 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 14110.xml