Fracture of a polyimide–elastimide posterior chamber intraocular lens following blunt orbital trauma. Issue 5 (September 2020)
- Record Type:
- Journal Article
- Title:
- Fracture of a polyimide–elastimide posterior chamber intraocular lens following blunt orbital trauma. Issue 5 (September 2020)
- Main Title:
- Fracture of a polyimide–elastimide posterior chamber intraocular lens following blunt orbital trauma
- Authors:
- Friehmann, Asaf
Sharon, Tal
Ophir, Shira Sheen
Assia, Ehud I - Abstract:
- Purpose: Fracture of a three-piece polyimide–elastimide intraocular lens at the optic–haptic junction following blunt trauma is a possible complication after cataract surgery. We report a case of an isolated posterior chamber intraocular lens fracture caused by direct ocular blunt trauma. Case: A 51-year-old patient underwent blunt orbital trauma caused by a raw chicken egg that was thrown at him, as he was walking. The posterior chamber intraocular lens was found in the anterior chamber, with a broken optic–haptic junction. The intraocular lens was exchanged, followed by iris fixation and pupilloplasty, with satisfactory postoperative anatomical and optical outcomes. Discussion: The three-piece polyimide–elastimide intraocular lens was fractured at the optic–haptic junction. Although the fractured intraocular lens surface had a regular appearance, it has been previously reported that the polyimide haptic's durability is probably inferior to that of polymethylmethacrylate. Therefore, it is highly susceptible to shear stress induced by a blunt trauma. Conclusion: To our knowledge, this is an uncommon report of an implanted posterior chamber polyimide–elastimide intraocular lens fracture following blunt orbital trauma. Intraocular lens exchange and fixation had successful optical and anatomical results.
- Is Part Of:
- European journal of ophthalmology. Volume 30:Issue 5(2020)
- Journal:
- European journal of ophthalmology
- Issue:
- Volume 30:Issue 5(2020)
- Issue Display:
- Volume 30, Issue 5 (2020)
- Year:
- 2020
- Volume:
- 30
- Issue:
- 5
- Issue Sort Value:
- 2020-0030-0005-0000
- Page Start:
- NP79
- Page End:
- NP81
- Publication Date:
- 2020-09
- Subjects:
- Posterior chamber intraocular lens breakage -- blunt trauma -- intraocular lens fixation -- intraocular lens exchange
Ophthalmology -- Periodicals
Eye -- Diseases -- Periodicals
617.7005 - Journal URLs:
- http://www.uk.sagepub.com/home.nav ↗
http://www.eur-j-ophthalmol.com/Home/Index ↗
http://journals.sagepub.com/home/ejo ↗ - DOI:
- 10.1177/1120672119838125 ↗
- Languages:
- English
- ISSNs:
- 1120-6721
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 14020.xml