Agreement and repeatability of four different devices to measure non-invasive tear breakup time (NIBUT). Issue 5 (October 2020)
- Record Type:
- Journal Article
- Title:
- Agreement and repeatability of four different devices to measure non-invasive tear breakup time (NIBUT). Issue 5 (October 2020)
- Main Title:
- Agreement and repeatability of four different devices to measure non-invasive tear breakup time (NIBUT)
- Authors:
- Bandlitz, Stefan
Peter, Barbara
Pflugi, Tanja
Jaeger, Kai
Anwar, Aaisha
Bikhu, Paramdeep
Nosch, Daniela S.
Wolffsohn, James S. - Abstract:
- Abstract: Purpose: Since tear film stability can be affected by fluorescein, the Dry Eye Workshop (DEWSII) recommended non-invasive measurement of tear breakup time (NIBUT). The aim of this study was to investigate the agreement and repeatability of four different instruments in the measurement of NIBUT. Methods: 72 participants (mean 24.2 ± 3.6 years) were recruited for this multi-centre, cross-sectional study. NIBUT was measured three times from one eye using each of the instruments in randomized order on two separate sessions during a day, separated by at least 2 h. NIBUT was performed at three sites (Switzerland, Germany and UK) using three subjective instruments, Tearscope Plus (Keeler, Windsor, UK) (TS), Polaris (bon Optic, Lübeck, Germany) (POL), EasyTear Viewplus (Easytear, Rovereto, Italy) (ET) and the objective Keratograph 5 M (Oculus Optikgeräte GmbH, Wetzlar, Germany) (KER). As the latter instrument only analyses for 24 s, all data was capped at this value. Results: NIBUT measurements (average of both sessions) between the four instruments were not statistically significantly different: TS (median 10.4, range 2.0–24.0 s), POL (10.1, 1.0–24.0 s), ET (10.6, 1.0–24.0 s) and KER (11.1, 2.6–24.0 s) (p = 0.949). The objective KER measures were on average (1.2 s ± 9.6 s, 95 % confidence interval) greater than the subjective evaluations of NIBUT with the other instruments (mean difference 0.4 s ± 7.7 s, 95 % confidence interval), resulting in a higher limits ofAbstract: Purpose: Since tear film stability can be affected by fluorescein, the Dry Eye Workshop (DEWSII) recommended non-invasive measurement of tear breakup time (NIBUT). The aim of this study was to investigate the agreement and repeatability of four different instruments in the measurement of NIBUT. Methods: 72 participants (mean 24.2 ± 3.6 years) were recruited for this multi-centre, cross-sectional study. NIBUT was measured three times from one eye using each of the instruments in randomized order on two separate sessions during a day, separated by at least 2 h. NIBUT was performed at three sites (Switzerland, Germany and UK) using three subjective instruments, Tearscope Plus (Keeler, Windsor, UK) (TS), Polaris (bon Optic, Lübeck, Germany) (POL), EasyTear Viewplus (Easytear, Rovereto, Italy) (ET) and the objective Keratograph 5 M (Oculus Optikgeräte GmbH, Wetzlar, Germany) (KER). As the latter instrument only analyses for 24 s, all data was capped at this value. Results: NIBUT measurements (average of both sessions) between the four instruments were not statistically significantly different: TS (median 10.4, range 2.0–24.0 s), POL (10.1, 1.0–24.0 s), ET (10.6, 1.0–24.0 s) and KER (11.1, 2.6–24.0 s) (p = 0.949). The objective KER measures were on average (1.2 s ± 9.6 s, 95 % confidence interval) greater than the subjective evaluations of NIBUT with the other instruments (mean difference 0.4 s ± 7.7 s, 95 % confidence interval), resulting in a higher limits of agreement. The slope was -0.08 to 0.11 indicating no bias in the difference between instruments with the magnitude of the NIBUT. Repeated measurements from the two sessions were not significantly different for TS (p = 0.584), POL (p = 0.549), ET (p = 0.701) or KER (p = 0.261). Conclusions: The four instruments evaluated for their measurement of tear stability were reasonably repeatable and give similar average results. … (more)
- Is Part Of:
- Contact lens & anterior eye. Volume 43:Issue 5(2020)
- Journal:
- Contact lens & anterior eye
- Issue:
- Volume 43:Issue 5(2020)
- Issue Display:
- Volume 43, Issue 5 (2020)
- Year:
- 2020
- Volume:
- 43
- Issue:
- 5
- Issue Sort Value:
- 2020-0043-0005-0000
- Page Start:
- 507
- Page End:
- 511
- Publication Date:
- 2020-10
- Subjects:
- Non-invasive breakup time -- Tear film stability -- Dry eye -- NIBUT -- Repeatability -- Tearscope -- Polaris -- Easy-tear -- Keratograph
Anterior segment (Eye) -- Periodicals
Contact lenses -- Periodicals
Segment antérieur (Œil) -- Périodiques
Lentilles de contact -- Périodiques
617.752305 - Journal URLs:
- http://firstsearch.oclc.org/journal=1367-0484;screen=info;ECOIP ↗
http://www.sciencedirect.com/science/journal/13670484 ↗
http://www.contactlensjournal.com/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/13670484 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/13670484 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.clae.2020.02.018 ↗
- Languages:
- English
- ISSNs:
- 1367-0484
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3424.971000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 14003.xml