Supra-annular sizing of transcatheter aortic valve prostheses in raphe-type bicuspid aortic valve disease: the LIRA method. (15th October 2020)
- Record Type:
- Journal Article
- Title:
- Supra-annular sizing of transcatheter aortic valve prostheses in raphe-type bicuspid aortic valve disease: the LIRA method. (15th October 2020)
- Main Title:
- Supra-annular sizing of transcatheter aortic valve prostheses in raphe-type bicuspid aortic valve disease: the LIRA method
- Authors:
- Iannopollo, Gianmarco
Romano, Vittorio
Buzzatti, Nicola
Ancona, Marco
Ferri, Luca
Russo, Filippo
Bellini, Barbara
Granada, Juan F.
Chieffo, Alaide
Montorfano, Matteo - Abstract:
- Abstract: Background: Recent evidence shows that THV prostheses anchoring occurs at the raphe-level, known as LIRA plane, in raphe-type bicuspid aortic valve (BAV) disease. The purpose of this study was to evaluate the application of a novel supra-annular sizing method, known as Level of Implantation at the RAphe (LIRA) method, to optimize transcatheter heart valve (THV) prosthesis sizing in raphe-type BAV disease. Methods and results: The LIRA method was applied to all consecutive patients with raphe-type BAV disease between November 2018 to January 2020 in our centre. THV prostheses were sized on the basis of baseline CT scan perimeters at the LIRA plane and at the virtual basal ring. In case of discrepancy between the two plane measurements, the plane with the smallest perimeter was considered the reference for prosthesis sizing. Post-procedural device success, defined according to Valve Academic Research Consortium–2 (VARC-2) criteria, was evaluated in the overall cohort. 20 patients (mean patient age 81 ± 5.4 years, 70% males) were identified as having a raphe-type BAV disease at pre-procedural CT scans and were implanted with different types of THV prostheses. The LIRA plane method appeared to be highly successful (100% VARC-2 device success) with no procedural mortality, no valve migration, no moderate-severe paravalvular leak and low transprosthetic gradient (residual mean gradient of 8.2 ± 2.9 mm Hg). Conclusions: Supra-annular sizing according to the LIRA methodAbstract: Background: Recent evidence shows that THV prostheses anchoring occurs at the raphe-level, known as LIRA plane, in raphe-type bicuspid aortic valve (BAV) disease. The purpose of this study was to evaluate the application of a novel supra-annular sizing method, known as Level of Implantation at the RAphe (LIRA) method, to optimize transcatheter heart valve (THV) prosthesis sizing in raphe-type BAV disease. Methods and results: The LIRA method was applied to all consecutive patients with raphe-type BAV disease between November 2018 to January 2020 in our centre. THV prostheses were sized on the basis of baseline CT scan perimeters at the LIRA plane and at the virtual basal ring. In case of discrepancy between the two plane measurements, the plane with the smallest perimeter was considered the reference for prosthesis sizing. Post-procedural device success, defined according to Valve Academic Research Consortium–2 (VARC-2) criteria, was evaluated in the overall cohort. 20 patients (mean patient age 81 ± 5.4 years, 70% males) were identified as having a raphe-type BAV disease at pre-procedural CT scans and were implanted with different types of THV prostheses. The LIRA plane method appeared to be highly successful (100% VARC-2 device success) with no procedural mortality, no valve migration, no moderate-severe paravalvular leak and low transprosthetic gradient (residual mean gradient of 8.2 ± 2.9 mm Hg). Conclusions: Supra-annular sizing according to the LIRA method appeared to be safe with a high device success. The application of the LIRA method might optimize THV prosthesis sizing in patients with raphe-type BAV disease. Highlights: THV prostheses anchoring occurs at the LIRA plane in raphe-type BAV disease. Supra-annular prosthesis sizing is not standardized in BAV disease. The LIRA sizing method appeared to be safe with a high device success. The LIRA method might optimize prosthesis sizing in raphe-type BAV disease. … (more)
- Is Part Of:
- International journal of cardiology. Volume 317(2020)
- Journal:
- International journal of cardiology
- Issue:
- Volume 317(2020)
- Issue Display:
- Volume 317, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 317
- Issue:
- 2020
- Issue Sort Value:
- 2020-0317-2020-0000
- Page Start:
- 144
- Page End:
- 151
- Publication Date:
- 2020-10-15
- Subjects:
- Aortic stenosis -- TAVI -- Bicuspid aortic valve -- MSCT
BAV bicuspid aortic valve -- ICD intercommissural distance -- MDCT multi-detector computed tomography -- LIRA Level of Implantation at the RAphe -- PVL paravalvular leak -- SA supra-annular -- SAVR surgical aortic valve replacement -- STS Society of Thoracic Surgeons -- TAVI transcatheter aortic valve replacement -- THV transcatheter heart valve -- VBR virtual basal ring
Cardiology -- Periodicals
Electronic journals
616.12 - Journal URLs:
- http://www.clinicalkey.com/dura/browse/journalIssue/01675273 ↗
http://www.sciencedirect.com/science/journal/01675273 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ijcard.2020.05.076 ↗
- Languages:
- English
- ISSNs:
- 0167-5273
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.158000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: