Performance characteristics of an instrument-free point-of-care CD4 test (VISITECT®CD4) for use in resource-limited settings. Issue 9 (September 2020)
- Record Type:
- Journal Article
- Title:
- Performance characteristics of an instrument-free point-of-care CD4 test (VISITECT®CD4) for use in resource-limited settings. Issue 9 (September 2020)
- Main Title:
- Performance characteristics of an instrument-free point-of-care CD4 test (VISITECT®CD4) for use in resource-limited settings
- Authors:
- Vidhyavathi, Vairamohan
Iqbal, Hussain Syed
Kanthamani, Kannaiyan
Solomon, Sunil Suhas
Nandagopal, Paneerselvam
Luchters, Stanley
Pham, Minh D.
Forbes, James
Crowe, Suzanne M.
McMurran, Christopher
Anderson, David A.
Balakrishnan, Pachamuthu - Abstract:
- Objective: CD4+ T lymphocyte count remains the most common biomarker of immune status and disease progression in human immunodeficiency virus (HIV)-positive individuals. VISITECT®CD4 is an instrument-free, low-cost point-of-care CD4 test with a cut-off of 350 CD4 cells/μL. This study aimed to evaluate VISITECT®CD4 test's diagnostic accuracy. Methods: Two hundred HIV-positive patients attending a tertiary HIV centre in South India were recruited. Patients provided venous blood for reference and VISITECT®CD4 tests. An additional finger-prick blood sample was obtained for VISITECT®CD4. VISITECT®CD4's diagnostic performance in identifying individuals with CD4 counts ≤350 cells/μL was assessed by calculating sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) taking flow cytometry as the reference. Results: The overall agreement between VISITECT®CD4 and flow cytometry was 89.5% using venous blood and 81.5% using finger-prick blood. VISITECT®CD4 showed better performance using venous blood [sensitivity: 96.6% (95% confidence interval: 92.1%–98.9%), specificity: 70.9% (57.1%–82.4%), PPV: 89.7% (83.9%–94.0%) and NPV: 88.6% (75.4%–96.2%)] than using finger-prick blood [sensitivity: 84.8% (77.9%–90.2%), specificity: 72.7% (59.0%–83.9%), PPV: 89.1% (82.7%–93.8%) and NPV: 64.5% (51.3%–76.3%)]. Conclusion: VISITECT®CD4 performed well using venous blood, demonstrating its potential utility in decentralization of CD4 testing services inObjective: CD4+ T lymphocyte count remains the most common biomarker of immune status and disease progression in human immunodeficiency virus (HIV)-positive individuals. VISITECT®CD4 is an instrument-free, low-cost point-of-care CD4 test with a cut-off of 350 CD4 cells/μL. This study aimed to evaluate VISITECT®CD4 test's diagnostic accuracy. Methods: Two hundred HIV-positive patients attending a tertiary HIV centre in South India were recruited. Patients provided venous blood for reference and VISITECT®CD4 tests. An additional finger-prick blood sample was obtained for VISITECT®CD4. VISITECT®CD4's diagnostic performance in identifying individuals with CD4 counts ≤350 cells/μL was assessed by calculating sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) taking flow cytometry as the reference. Results: The overall agreement between VISITECT®CD4 and flow cytometry was 89.5% using venous blood and 81.5% using finger-prick blood. VISITECT®CD4 showed better performance using venous blood [sensitivity: 96.6% (95% confidence interval: 92.1%–98.9%), specificity: 70.9% (57.1%–82.4%), PPV: 89.7% (83.9%–94.0%) and NPV: 88.6% (75.4%–96.2%)] than using finger-prick blood [sensitivity: 84.8% (77.9%–90.2%), specificity: 72.7% (59.0%–83.9%), PPV: 89.1% (82.7%–93.8%) and NPV: 64.5% (51.3%–76.3%)]. Conclusion: VISITECT®CD4 performed well using venous blood, demonstrating its potential utility in decentralization of CD4 testing services in resource-constrained settings. … (more)
- Is Part Of:
- Journal of international medical research. Volume 48:Issue 9(2020)
- Journal:
- Journal of international medical research
- Issue:
- Volume 48:Issue 9(2020)
- Issue Display:
- Volume 48, Issue 9 (2020)
- Year:
- 2020
- Volume:
- 48
- Issue:
- 9
- Issue Sort Value:
- 2020-0048-0009-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-09
- Subjects:
- VISITECT®CD4 test -- point-of-care diagnostics -- CD4+ count -- HIV monitoring assay -- HIV/AIDS -- resource-limited settings
Medicine -- Periodicals
Pharmacology -- Periodicals
610.5 - Journal URLs:
- http://imr.sagepub.com/ ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0300060520955028 ↗
- Languages:
- English
- ISSNs:
- 0300-0605
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13961.xml