Combined two-centre experience of single-entry telescopic rods identifies characteristic modes of failure. (August 2020)
- Record Type:
- Journal Article
- Title:
- Combined two-centre experience of single-entry telescopic rods identifies characteristic modes of failure. (August 2020)
- Main Title:
- Combined two-centre experience of single-entry telescopic rods identifies characteristic modes of failure
- Authors:
- Cox, India
Al Mouazzen, Louay
Bleibleh, Sabri
Moldovan, Radu
Bintcliffe, Fiona
Bache, C. Edward
Thomas, Simon - Abstract:
- Abstract : Aims: The Fassier Duval (FD) rod is a third-generation telescopic implant for children with osteogenesis imperfecta (OI). Threaded fixation enables proximal insertion without opening the knee or ankle joint. We have reviewed our combined two-centre experience with this implant. Methods: In total, 34 children with a mean age of five years (1 to 14) with severe OI have undergone rodding of 72 lower limb long bones (27 tibial, 45 femoral) for recurrent fractures with progressive deformity despite optimized bone health and bisphosphonate therapy. Data were collected prospectively, with 1.5 to 11 years follow-up. Results: A total of 24 patients (33%) required exchange of implants (14 femora and ten tibiae) including 11 rods bending with refracture. Four (5%) required reoperation with implant retention. Loss of proximal fixation in the femur and distal fixation in the tibia were common. Four patients developed coxa vara requiring surgical correction. In total, 13 patients experienced further fractures without rod bending; eight required implant revision. There was one deep infection. The five-year survival rate, with rod revision as the endpoint, was 63% (95% confidence interval (CI) 44% to 77%) for femoral rods, with a mean age at implantation of 4.8 years (1.3 to 14.8), and 64% (95% CI 36% to 82%) for tibial rods, with a mean age at implantation of 5.2 years (2.0 to 13.8). Conclusion: FD rods are easier to implant but do not improve on the revision rates reported forAbstract : Aims: The Fassier Duval (FD) rod is a third-generation telescopic implant for children with osteogenesis imperfecta (OI). Threaded fixation enables proximal insertion without opening the knee or ankle joint. We have reviewed our combined two-centre experience with this implant. Methods: In total, 34 children with a mean age of five years (1 to 14) with severe OI have undergone rodding of 72 lower limb long bones (27 tibial, 45 femoral) for recurrent fractures with progressive deformity despite optimized bone health and bisphosphonate therapy. Data were collected prospectively, with 1.5 to 11 years follow-up. Results: A total of 24 patients (33%) required exchange of implants (14 femora and ten tibiae) including 11 rods bending with refracture. Four (5%) required reoperation with implant retention. Loss of proximal fixation in the femur and distal fixation in the tibia were common. Four patients developed coxa vara requiring surgical correction. In total, 13 patients experienced further fractures without rod bending; eight required implant revision. There was one deep infection. The five-year survival rate, with rod revision as the endpoint, was 63% (95% confidence interval (CI) 44% to 77%) for femoral rods, with a mean age at implantation of 4.8 years (1.3 to 14.8), and 64% (95% CI 36% to 82%) for tibial rods, with a mean age at implantation of 5.2 years (2.0 to 13.8). Conclusion: FD rods are easier to implant but do not improve on the revision rates reported for second generation T-piece rods. Proximal femoral fixation is problematic in younger children with a partially ossified greater trochanter. Distal tibial fixation typically fails after two years. Future generation implants should address proximal femoral and distal tibial fixation to avoid the majority of complications in this series. Cite this article: Bone Joint J 2020;102-B(8):1048–1055. … (more)
- Is Part Of:
- Bone & joint journal. Volume 102B:Number 8(2020)
- Journal:
- Bone & joint journal
- Issue:
- Volume 102B:Number 8(2020)
- Issue Display:
- Volume 102, Issue 8 (2020)
- Year:
- 2020
- Volume:
- 102
- Issue:
- 8
- Issue Sort Value:
- 2020-0102-0008-0000
- Page Start:
- 1048
- Page End:
- 1055
- Publication Date:
- 2020-08
- Subjects:
- Paediatrics -- Osteogenesis imperfecta -- Telescopic rods
Bones -- Surgery -- Periodicals
Joints -- Surgery -- Periodicals
Orthopedic surgery -- Periodicals
617.47005 - Journal URLs:
- http://www.bjj.boneandjoint.org.uk/ ↗
- DOI:
- 10.1302/0301-620X.102B8.BJJ-2020-0131.R1 ↗
- Languages:
- English
- ISSNs:
- 2049-4394
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library STI - ELD Digital store
- Ingest File:
- 13849.xml