Effect of supporting implants inclination on stability of fixed partial denture: A finite element study. (October 2020)
- Record Type:
- Journal Article
- Title:
- Effect of supporting implants inclination on stability of fixed partial denture: A finite element study. (October 2020)
- Main Title:
- Effect of supporting implants inclination on stability of fixed partial denture: A finite element study
- Authors:
- Boukhlif, Amel
Merdji, Ali
Roy, Sandipan
Alkhaldi, Hashem
Abu-Alshaikh, Ibrahim
Della, Nourddine
Cristache, Corina Marilena
Hillstrom, Rajshree - Abstract:
- The aim of this finite element study was to analyze effect of supporting implants inclination on stress distribution in the bone for a four-unit fixed partial denture. A three-dimensional finite element model of mandibular molar section of the bone to receive implants was constructed. Three implant-supported fixed partial dentures, with null, moderate and wide tilting, of 0°, 15° and 30° implant inclinations, respectively, were modeled. A mechanical load of 10 MPa was applied in coronal–apical direction on bridge framework at the regions of crowns positions. The finite element analysis was performed, and von Mises stress levels were calculated. Peak stress concentration in the cortical bone was observed mostly around the implant necks, in inter-implants line. There was favorable stress distribution during loading, with peak stress being 90.04 MPa for 0°, which decreased to 54.33 MPa for 15° and 46.36 MPa for 30° inclination. The supporting implants inclination in fixed partial denture plays an important role in stress distribution and may be helpful in preventing bone loss and implant failure. This phenomenon is likely to be more pronounced in bones of poor quality. Within the limitation of this study, it seems that the inclination of implants in fixed partial denture has a favorable effect on stress distribution pattern values around the supporting implants.
- Is Part Of:
- Proceedings of the Institution of Mechanical Engineers. Volume 234:Number 10(2020)
- Journal:
- Proceedings of the Institution of Mechanical Engineers
- Issue:
- Volume 234:Number 10(2020)
- Issue Display:
- Volume 234, Issue 10 (2020)
- Year:
- 2020
- Volume:
- 234
- Issue:
- 10
- Issue Sort Value:
- 2020-0234-0010-0000
- Page Start:
- 1162
- Page End:
- 1171
- Publication Date:
- 2020-10
- Subjects:
- Implant inclination -- stress -- finite elements analysis -- dental biomechanics -- bone
Biomedical engineering -- Periodicals
Medical instruments and apparatus -- Periodicals
610.28 - Journal URLs:
- http://pih.sagepub.com/ ↗
http://journals.pepublishing.com/content/119779 ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0954411920944109 ↗
- Languages:
- English
- ISSNs:
- 0954-4119
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13592.xml