Study on the bending behavior of biodegradable metal cerebral vascular stents using finite element analysis. (17th July 2020)
- Record Type:
- Journal Article
- Title:
- Study on the bending behavior of biodegradable metal cerebral vascular stents using finite element analysis. (17th July 2020)
- Main Title:
- Study on the bending behavior of biodegradable metal cerebral vascular stents using finite element analysis
- Authors:
- Shi, Weiliang
Li, Hongxia
Zhu, Tingzhun
Jin, Yifei
Wang, Hairui
Yang, Jianbing
Zhao, Danyang - Abstract:
- Graphical abstract: Abstract: Excellent bending behavior is evaluated as the primary factor during the design of biodegradable metal cerebral vascular stents (BMCVSs), which enables vascular stents to be successfully delivered to the targeted location and avoids unnecessary damage to blood vessels. Unfortunately, this bending behavior has been barely investigated which limits the design of BMCVSs with optimal structures. Herein, six BMCVSs were designed and their bending process were simulated using finite element analysis (FEA) . Then, the effects of the stent bridge connection type and structure on the bending behavior were systematically analyzed and an universal mathematical model was further established, in which the influence of the structure parameters of the stent bridge on the flexibility of stents was considered. After that, the bending mechanism of the high-stress zone of the bridge was investigated. Finally, the causes and effects of the self-contacting phenomenon as well as the inner-stent protrusion phenomenon in the bending state were analyzed theoretically, and corresponding solutions were proposed to optimize the design of stents. The numerical results show that the stents with the dislocation-line W-shaped unit have better flexibility than the other stents. The flexibility is positively correlated to the cube of the length of linear part and to the square of the curvature of curved part. The self-contacting phenomenon of the bridge during bending canGraphical abstract: Abstract: Excellent bending behavior is evaluated as the primary factor during the design of biodegradable metal cerebral vascular stents (BMCVSs), which enables vascular stents to be successfully delivered to the targeted location and avoids unnecessary damage to blood vessels. Unfortunately, this bending behavior has been barely investigated which limits the design of BMCVSs with optimal structures. Herein, six BMCVSs were designed and their bending process were simulated using finite element analysis (FEA) . Then, the effects of the stent bridge connection type and structure on the bending behavior were systematically analyzed and an universal mathematical model was further established, in which the influence of the structure parameters of the stent bridge on the flexibility of stents was considered. After that, the bending mechanism of the high-stress zone of the bridge was investigated. Finally, the causes and effects of the self-contacting phenomenon as well as the inner-stent protrusion phenomenon in the bending state were analyzed theoretically, and corresponding solutions were proposed to optimize the design of stents. The numerical results show that the stents with the dislocation-line W-shaped unit have better flexibility than the other stents. The flexibility is positively correlated to the cube of the length of linear part and to the square of the curvature of curved part. The self-contacting phenomenon of the bridge during bending can constrain the formation of inner-stent protrusion, which can eliminate the negative effects of the implanted stents on the hemodynamics in blood vessels. This study is expected to provide practical guidance for the structural design of BMCVSs for clinical applications. … (more)
- Is Part Of:
- Journal of biomechanics. Volume 108(2020)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 108(2020)
- Issue Display:
- Volume 108, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 108
- Issue:
- 2020
- Issue Sort Value:
- 2020-0108-2020-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-07-17
- Subjects:
- Cerebral vascular stent -- Biodegradable -- Bending behavior -- Flexibility -- Finite element analysis
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2020.109856 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13587.xml