Determination of a consensus scaling factor to convert a Co-60-based alanine dose reading to yield the dose delivered in a high energy electron beam. (June 2020)
- Record Type:
- Journal Article
- Title:
- Determination of a consensus scaling factor to convert a Co-60-based alanine dose reading to yield the dose delivered in a high energy electron beam. (June 2020)
- Main Title:
- Determination of a consensus scaling factor to convert a Co-60-based alanine dose reading to yield the dose delivered in a high energy electron beam
- Authors:
- McEwen, Malcolm
Miller, Arne
Pazos, Ileana
Sharpe, Peter - Abstract:
- Abstract: An analysis of the published literature has been carried out to determine a consensus value for the scaling factor required to convert a dose determination using alanine based on a Co-60 calibration to yield the dose delivered in a high energy electron beam (≥6 MeV incident). Published data from the last 15 years were reviewed and combined to give a single, average, scaling factor. In carrying out this analysis, no attempt was made to deconstruct each experiment. The final results of each investigation were used as a coherent ensemble of independent determinations, irrespective of method or alanine dosimeter supplier. Experimental and Monte Calculations were considered separately and only the experimental data was used to determine the scaling factor. The results from six investigations were found to be consistent within their uncertainties and an overall conversion factor of 1.014 (standard uncertainty = 0.5%) was obtained. To be explicit, alanine dose readings determined from a Co-60 alanine calibration should be multiplied by this factor to yield the dose delivered in an electron beam. This scaling factor represents the consensus recommendations of the National Research Council, the National Institute for Standards and Technology, the Technical University of Denmark and the National Physical Laboratory for electron beams (≥6 MeV incident) used in radiation processing applications. By adopting a common factor, it is hoped to improve the consistency of dosimetryAbstract: An analysis of the published literature has been carried out to determine a consensus value for the scaling factor required to convert a dose determination using alanine based on a Co-60 calibration to yield the dose delivered in a high energy electron beam (≥6 MeV incident). Published data from the last 15 years were reviewed and combined to give a single, average, scaling factor. In carrying out this analysis, no attempt was made to deconstruct each experiment. The final results of each investigation were used as a coherent ensemble of independent determinations, irrespective of method or alanine dosimeter supplier. Experimental and Monte Calculations were considered separately and only the experimental data was used to determine the scaling factor. The results from six investigations were found to be consistent within their uncertainties and an overall conversion factor of 1.014 (standard uncertainty = 0.5%) was obtained. To be explicit, alanine dose readings determined from a Co-60 alanine calibration should be multiplied by this factor to yield the dose delivered in an electron beam. This scaling factor represents the consensus recommendations of the National Research Council, the National Institute for Standards and Technology, the Technical University of Denmark and the National Physical Laboratory for electron beams (≥6 MeV incident) used in radiation processing applications. By adopting a common factor, it is hoped to improve the consistency of dosimetry at industrial dose levels using high-energy electron beams. Highlights: Recommendation by the lead laboratories providing alanine calibration services. Review of literature from last 15 years on scaling from Co-60 to electron beams. Analysis of experimental and calculated values. … (more)
- Is Part Of:
- Radiation physics and chemistry. Volume 171(2020:Jun.)
- Journal:
- Radiation physics and chemistry
- Issue:
- Volume 171(2020:Jun.)
- Issue Display:
- Volume 171 (2020)
- Year:
- 2020
- Volume:
- 171
- Issue Sort Value:
- 2020-0171-0000-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-06
- Subjects:
- Alanine -- High-dose dosimetry -- High-energy electron beams
Radiation chemistry -- Periodicals
Radiometry -- Periodicals
Radiation -- Periodicals
Chimie sous rayonnement -- Périodiques
539.2 - Journal URLs:
- http://www.sciencedirect.com/science/journal/0969806X ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiation-physics-and-chemistry/ ↗ - DOI:
- 10.1016/j.radphyschem.2019.108673 ↗
- Languages:
- English
- ISSNs:
- 0969-806X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7227.984000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13497.xml