The impact of gonad shielding in anteroposterior (AP) pelvis projections in an adult: A phantom study utilising digital radiography (DR). Issue 3 (August 2020)
- Record Type:
- Journal Article
- Title:
- The impact of gonad shielding in anteroposterior (AP) pelvis projections in an adult: A phantom study utilising digital radiography (DR). Issue 3 (August 2020)
- Main Title:
- The impact of gonad shielding in anteroposterior (AP) pelvis projections in an adult: A phantom study utilising digital radiography (DR)
- Authors:
- Davies, B.H.
Manning-Stanley, A.S.
Hughes, V.J.
Ward, A.J. - Abstract:
- Abstract: Introduction: Positioning relative to the lateral automatic exposure control (AEC) chambers (cranial/caudal orientation) optimises dose and image quality in pelvic radiography. In the cranial orientation introducing gonad shielding (GS) in females may increase radiation dose. The aim of this study was to fully optimise the combination of pelvis orientation and use of GS in both male and females. Methods: An anthropomorphic pelvis phantom was exposed, with dose area product (DAP) recorded, in both orientations without GS and four conditions with GS: cranial orientation (female/male), caudal orientation (female/male). A 4 cm × 4 cm grid incorporating thirteen positions for the GS resulted in 52 experimental settings. Blind image quality assessment, utilising a modified scale, was undertaken by two experienced observers. Results: Comparing no GS (caudal orientation) to female GS, no significant change in DAP was seen (3.97 v 4.03 dGy*cm 2 ; Mann–Whitney p = 0.060). Comparing no GS (cranial orientation) to male GS no significant change in DAP was seen (8.66 v 8.77 dGy*cm 2 ; Mann–Whitney; p = 0.210). DAP increased significantly with introduction of female GS in the cranial orientation (23%: 8.66 v 10.65 dGy*cm 2, Mann–Whitney; p < 0.001) and male GS in the caudal orientation (22.8%: 3.97 v 4.87 dGy*cm 2, Mann–Whitney; p < 0.001). Significantly higher repeat rates (Chi-squared test; p < 0.001) were seen for GS in female (85–100%) compared to male (30.8%). Conclusion:Abstract: Introduction: Positioning relative to the lateral automatic exposure control (AEC) chambers (cranial/caudal orientation) optimises dose and image quality in pelvic radiography. In the cranial orientation introducing gonad shielding (GS) in females may increase radiation dose. The aim of this study was to fully optimise the combination of pelvis orientation and use of GS in both male and females. Methods: An anthropomorphic pelvis phantom was exposed, with dose area product (DAP) recorded, in both orientations without GS and four conditions with GS: cranial orientation (female/male), caudal orientation (female/male). A 4 cm × 4 cm grid incorporating thirteen positions for the GS resulted in 52 experimental settings. Blind image quality assessment, utilising a modified scale, was undertaken by two experienced observers. Results: Comparing no GS (caudal orientation) to female GS, no significant change in DAP was seen (3.97 v 4.03 dGy*cm 2 ; Mann–Whitney p = 0.060). Comparing no GS (cranial orientation) to male GS no significant change in DAP was seen (8.66 v 8.77 dGy*cm 2 ; Mann–Whitney; p = 0.210). DAP increased significantly with introduction of female GS in the cranial orientation (23%: 8.66 v 10.65 dGy*cm 2, Mann–Whitney; p < 0.001) and male GS in the caudal orientation (22.8%: 3.97 v 4.87 dGy*cm 2, Mann–Whitney; p < 0.001). Significantly higher repeat rates (Chi-squared test; p < 0.001) were seen for GS in female (85–100%) compared to male (30.8%). Conclusion: The use of gonad shielding can increase DAP and lead to repeats being required, with more required for female GS usage, suggesting the utility of GS for pelvis examinations is questionable. Implications for practice: Optimisation of radiation dose in pelvic radiographic examinations utilising AEC terminated exposures requires consideration of AEC chamber position and GS usage . … (more)
- Is Part Of:
- Radiography. Volume 26:Issue 3(2020)
- Journal:
- Radiography
- Issue:
- Volume 26:Issue 3(2020)
- Issue Display:
- Volume 26, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 26
- Issue:
- 3
- Issue Sort Value:
- 2020-0026-0003-0000
- Page Start:
- 240
- Page End:
- 247
- Publication Date:
- 2020-08
- Subjects:
- Dose optimisation -- Gonad shielding -- AEC -- Pelvis -- Phantom
Diagnostic imaging -- Periodicals
Radiotherapy -- Periodicals
Cancer -- Radiotherapy -- Periodicals
Diagnostic Imaging -- Periodicals
Neoplasms -- Periodicals
Radiotherapy -- Periodicals
Imagerie pour le diagnostic -- Périodiques
Radiothérapie -- Périodiques
Cancer -- Radiothérapie -- Périodiques
Electronic journals
616.0757 - Journal URLs:
- http://www.sciencedirect.com/science/journal/10788174 ↗
http://www.radiographyonline.com/ ↗
http://www.harcourt-international.com/journals ↗
http://www.idealibrary.com/links/toc/radi/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/10788174 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/10788174 ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiography/ ↗ - DOI:
- 10.1016/j.radi.2020.01.007 ↗
- Languages:
- English
- ISSNs:
- 1078-8174
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7237.001000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 13435.xml