The calibration and deployment of a low-cost methane sensor. (1st June 2020)
- Record Type:
- Journal Article
- Title:
- The calibration and deployment of a low-cost methane sensor. (1st June 2020)
- Main Title:
- The calibration and deployment of a low-cost methane sensor
- Authors:
- Riddick, Stuart N.
Mauzerall, Denise L.
Celia, Michael
Allen, Grant
Pitt, Joseph
Kang, Mary
Riddick, John C. - Abstract:
- Abstract: Since 1850 the atmospheric mixing ratio of methane (CH4 ), a potent greenhouse gas, has doubled. This increase is directly linked to an escalation in emissions from anthropogenic sources. An inexpensive means to identify and monitor CH4 emission sources and evaluate the efficacy of mitigation strategies is essential. However, sourcing reliable, low-cost, easy-to-calibrate sensors that are fit for purpose is challenging. A recent study showed that CH4 mixing ratio data from a low-power, low-cost CH4 sensor (Figaro TGS2600) agreed well with CH4 mixing ratios measured by a high precision sensor at mixing ratios between 1.85 ppm and 2 ppm. To investigate, as a proof of concept, if this low-cost sensor could be used to measure typical ambient CH4 mixing ratios, we operated a TGS2600 in conjunction with a Los Gatos Ultra-portable Greenhouse Gas Analyzer (UGGA) in controlled laboratory conditions. We then explored the sensor's long-term reliability by deploying the TGS2600 near an onshore gas terminal to calculate emissions from May to July 2018. Our initial studies showed that previously published linear algorithms could not convert TGS2600 output to CH4 mixing ratios measured by the UGGA. However, we derived a non-linear empirical relationship that could be used to reliably convert the output of a TGS2600 unit to CH4 mixing ratios over a range of 1.85–5.85 ppm that agree to a high-precision instrument output to ±0.01 ppm. Our study showed that the TGS2600 could be usedAbstract: Since 1850 the atmospheric mixing ratio of methane (CH4 ), a potent greenhouse gas, has doubled. This increase is directly linked to an escalation in emissions from anthropogenic sources. An inexpensive means to identify and monitor CH4 emission sources and evaluate the efficacy of mitigation strategies is essential. However, sourcing reliable, low-cost, easy-to-calibrate sensors that are fit for purpose is challenging. A recent study showed that CH4 mixing ratio data from a low-power, low-cost CH4 sensor (Figaro TGS2600) agreed well with CH4 mixing ratios measured by a high precision sensor at mixing ratios between 1.85 ppm and 2 ppm. To investigate, as a proof of concept, if this low-cost sensor could be used to measure typical ambient CH4 mixing ratios, we operated a TGS2600 in conjunction with a Los Gatos Ultra-portable Greenhouse Gas Analyzer (UGGA) in controlled laboratory conditions. We then explored the sensor's long-term reliability by deploying the TGS2600 near an onshore gas terminal to calculate emissions from May to July 2018. Our initial studies showed that previously published linear algorithms could not convert TGS2600 output to CH4 mixing ratios measured by the UGGA. However, we derived a non-linear empirical relationship that could be used to reliably convert the output of a TGS2600 unit to CH4 mixing ratios over a range of 1.85–5.85 ppm that agree to a high-precision instrument output to ±0.01 ppm. Our study showed that the TGS2600 could be used to continuously measure variability in CH4 mixing ratios from 1.82 to 5.40 ppm for three months downwind of the gas terminal. Using a simplified Gaussian Plume approach, these mixing ratios correspond to an emission flux range of 0–238 g CH4 s −1, with average emission of 9.6 g CH4 s −1 from the currently active North Terminal and 1.6 g CH4 s −1 from the decommissioned South Terminal. Our work here demonstrates the feasibility of utilizing a low-cost sensor to detect methane leakage at concentrations close to ambient background levels, as long as the device is routinely calibrated with an accurate reference instrument. Having a widely deployed network of such low-cost CH4 sensors would allow improved identification, monitoring and mitigation of a variety of CH4 emissions. Highlights: Low-cost, low-power methane sensor is investigated for use in a measurement network. Calibration shows a non-linear response to methane change between 2 and 6 ppm. Long-term deployment shows sensor is reliable and can be used to measure autonomously. Sensor can detect methane close to background but requires routine calibration. A network of low-cost sensors could help identification and mitigation of methane. … (more)
- Is Part Of:
- Atmospheric environment. Volume 230(2020)
- Journal:
- Atmospheric environment
- Issue:
- Volume 230(2020)
- Issue Display:
- Volume 230, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 230
- Issue:
- 2020
- Issue Sort Value:
- 2020-0230-2020-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-06-01
- Subjects:
- Methane -- Mixing ratio -- Measurement -- Inexpensive -- Calibration
Air -- Pollution -- Periodicals
Air -- Pollution -- Meteorological aspects -- Periodicals
551.51 - Journal URLs:
- http://www.sciencedirect.com/web-editions/journal/13522310 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.atmosenv.2020.117440 ↗
- Languages:
- English
- ISSNs:
- 1352-2310
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1767.120000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: