Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study. (May 2020)
- Record Type:
- Journal Article
- Title:
- Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study. (May 2020)
- Main Title:
- Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study
- Authors:
- Ranjbar, Sahar
Gholami, Somayeh
Longo, Francesco
Sharp, Ryan
Meigooni, Ali S. - Abstract:
- Abstract: Purpose: The aim of this work is to investigate the influence of an external magnetic field (MF) on The American Association of Physicists in Medicine (AAPM) No. 43 Report (TG-43) parameters for 192 Ir and 60 Co high dose rate (HDR) brachytherapy sources using Monte Carlo (MC) simulation methods. Materials and methods: We used the Geant4 toolkit (version 10.1. p01) to simulate the geometry of 192 Ir and 60 Co brachytherapy sources. AAPM TG-43 parameters (the radial dose function, g(r), and the anisotropy function, F (r, θ)) of both 192 Ir and 60 Co sources were calculated in the presence of a magnetic field with strengths of 1.5T, 3T, and 7T in the X, Y, and Z directions in a voxelized water phantom. Results: For the 192 Ir source, the calculated values g(r) and F (r, θ) remained nearly unaffected by the magnetic field for all investigated strengths. For the 60 Co source, the differences for the g(r) and F (r, θ) under the 1.5T, 3T, and 7T magnetic field strengths along the direction parallel with the MF were found to be an increase of up to 5%, 15%, and 33%, respectively. However, for the directions perpendicular with the magnetic field, there was a decrease of up to 3%, 6% and 15% under 1.5T, 3T and 7T strengths, respectively. Conclusion: Our results highlight the necessity of a Monte Carlo-based treatment planning system (TPS) if cobalt HDR treatments are performed under a magnetic field, especially for strengths greater than 1.5T. Highlights: For the 192 IrAbstract: Purpose: The aim of this work is to investigate the influence of an external magnetic field (MF) on The American Association of Physicists in Medicine (AAPM) No. 43 Report (TG-43) parameters for 192 Ir and 60 Co high dose rate (HDR) brachytherapy sources using Monte Carlo (MC) simulation methods. Materials and methods: We used the Geant4 toolkit (version 10.1. p01) to simulate the geometry of 192 Ir and 60 Co brachytherapy sources. AAPM TG-43 parameters (the radial dose function, g(r), and the anisotropy function, F (r, θ)) of both 192 Ir and 60 Co sources were calculated in the presence of a magnetic field with strengths of 1.5T, 3T, and 7T in the X, Y, and Z directions in a voxelized water phantom. Results: For the 192 Ir source, the calculated values g(r) and F (r, θ) remained nearly unaffected by the magnetic field for all investigated strengths. For the 60 Co source, the differences for the g(r) and F (r, θ) under the 1.5T, 3T, and 7T magnetic field strengths along the direction parallel with the MF were found to be an increase of up to 5%, 15%, and 33%, respectively. However, for the directions perpendicular with the magnetic field, there was a decrease of up to 3%, 6% and 15% under 1.5T, 3T and 7T strengths, respectively. Conclusion: Our results highlight the necessity of a Monte Carlo-based treatment planning system (TPS) if cobalt HDR treatments are performed under a magnetic field, especially for strengths greater than 1.5T. Highlights: For the 192 Ir source, the TG-43 parameters remained nearly unaffected by the magnetic field for all investigated strengths. For the 60 Co source, the TG-43 parameters can differ up to 33% under 7T magnetic field. The necessity of a Monte Carlo-based TPS was demonstrated if 60 Co HDR treatments are performed under a magnetic field. … (more)
- Is Part Of:
- Applied radiation and isotopes. Volume 159(2020:May)
- Journal:
- Applied radiation and isotopes
- Issue:
- Volume 159(2020:May)
- Issue Display:
- Volume 159 (2020)
- Year:
- 2020
- Volume:
- 159
- Issue Sort Value:
- 2020-0159-0000-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-05
- Subjects:
- HDR brachytherapy -- Geant4 simulation -- Magnetic fields -- 192Ir and 60Co sources
Radiology -- Periodicals
Radiation -- Industrial applications -- Periodicals
Nuclear chemistry -- Periodicals
Internet resource
Periodical
660.298 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09698043 ↗
http://catalog.hathitrust.org/api/volumes/oclc/27456684.html ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.apradiso.2020.109088 ↗
- Languages:
- English
- ISSNs:
- 0969-8043
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1576.565000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13382.xml