CoCrMo metal release in metal‐on‐highly crosslinked polyethylene hip implants. Issue 4 (19th August 2019)
- Record Type:
- Journal Article
- Title:
- CoCrMo metal release in metal‐on‐highly crosslinked polyethylene hip implants. Issue 4 (19th August 2019)
- Main Title:
- CoCrMo metal release in metal‐on‐highly crosslinked polyethylene hip implants
- Authors:
- Wang, Qiong
Eltit, Felipe
Garbuz, Donald
Duncan, Clive
Masri, Bassam
Greidanus, Nelson
Wang, Rizhi - Abstract:
- Abstract: Increasing cases of adverse local tissue reactions (ALTRs) associated with metal release have been observed in patients with metal‐on‐highly crosslinked polyethylene (MoP) hip implants, the most common design in total hip replacements. Studies have demonstrated the metal release from fretting corrosion at the head–neck junction, but rarely investigated tribocorrosion associated metal release at articulating surfaces in MoP hip implants. The objective of this study is to investigate both tribocorrosion at the articulating surfaces and fretting corrosion at the head–neck junction in CoCrMo femoral heads, as well as their association with metal species released in periprosthetic tissues and body fluids in MoP hip systems. Twenty‐three patients with ALTRs associated with MoP implants were included. Systematic analyses were performed on the wear damage in articulation, corrosion at the head–neck junction and their correlation with degradation products observed in synovial fluid, periprosthetic tissues, and serum. Results showed that tribocorrosion at the articulating surfaces contributed to the elevated concentration of both Co and Cr ions in serum, while fretting corrosion at the head–neck junction mainly released Co ions to serum. Both tribocorrosion at the articulating surfaces and fretting corrosion at the head–neck junction released particles rich in chromium and phosphate, the dominant particles found in synovial fluids and tissues. This study provides strongAbstract: Increasing cases of adverse local tissue reactions (ALTRs) associated with metal release have been observed in patients with metal‐on‐highly crosslinked polyethylene (MoP) hip implants, the most common design in total hip replacements. Studies have demonstrated the metal release from fretting corrosion at the head–neck junction, but rarely investigated tribocorrosion associated metal release at articulating surfaces in MoP hip implants. The objective of this study is to investigate both tribocorrosion at the articulating surfaces and fretting corrosion at the head–neck junction in CoCrMo femoral heads, as well as their association with metal species released in periprosthetic tissues and body fluids in MoP hip systems. Twenty‐three patients with ALTRs associated with MoP implants were included. Systematic analyses were performed on the wear damage in articulation, corrosion at the head–neck junction and their correlation with degradation products observed in synovial fluid, periprosthetic tissues, and serum. Results showed that tribocorrosion at the articulating surfaces contributed to the elevated concentration of both Co and Cr ions in serum, while fretting corrosion at the head–neck junction mainly released Co ions to serum. Both tribocorrosion at the articulating surfaces and fretting corrosion at the head–neck junction released particles rich in chromium and phosphate, the dominant particles found in synovial fluids and tissues. This study provides strong evidence that tribocorrosion at the articulating surfaces in MoP hip implants could result in significant metal release. This information should be taken into account when studying the mechanisms of ALTRs and developing strategies of preventing metal release in total hip replacements. … (more)
- Is Part Of:
- Journal of biomedical materials research. Volume 108:Issue 4(2020)
- Journal:
- Journal of biomedical materials research
- Issue:
- Volume 108:Issue 4(2020)
- Issue Display:
- Volume 108, Issue 4 (2020)
- Year:
- 2020
- Volume:
- 108
- Issue:
- 4
- Issue Sort Value:
- 2020-0108-0004-0000
- Page Start:
- 1213
- Page End:
- 1228
- Publication Date:
- 2019-08-19
- Subjects:
- adverse local tissue reactions -- CoCrMo alloys -- fretting corrosion -- hip implants -- metal release -- metal‐on‐polyethylene -- tribocorrosion
Biomedical materials -- Periodicals
610.28 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jbm.b.34470 ↗
- Languages:
- English
- ISSNs:
- 1552-4973
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.725000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13248.xml