Effect of blood viscosity on the performance of virtually wall-less venous cannulas. (July 2020)
- Record Type:
- Journal Article
- Title:
- Effect of blood viscosity on the performance of virtually wall-less venous cannulas. (July 2020)
- Main Title:
- Effect of blood viscosity on the performance of virtually wall-less venous cannulas
- Authors:
- Abdel-Sayed, Saad
Abdel-Sayed, Philippe
Berdaj, Denis
Ferrari, Enrico
Halbe, Maximilian
von Segesser, Ludwig Karl
Maisano, Francesco - Abstract:
- Aim: This study was designed to quantify the influence of blood as test medium compared to water in cannula bench performance assessment. Methods: An in vitro circuit was set-up with silicone tubing between two reservoirs. The test medium was pumped from the lower reservoir by centrifugal pump to the upper reservoir. The test-cannula was inserted in a silicone tube connected between the lower reservoir and the centrifugal pump. Flow rate and pump inlet-pressure were measured for wall-less versus thin-wall cannula using a centrifugal pump in a dynamic bench-test for an afterload of 40-60 mmHg using two media: blood 10 g/dL and 5.6 g/dL and water 0 g/dL. Results: The wall-less cannula showed significantly higher flows rates as compared to the thin-wall cannula (control), with both hemoglobin concentrations and water. Indeed, for a target volume of 200-250 mL of blood (Hg 10 g/dL) in the upper reservoir, the cannula outlet pressure (P) was −14 ± 14 mmHg versus −18 ± 11 mmHg for the wall-less and control respectively; the cannula outlet flow rate (Q) was 3.91 ± 0.41 versus 3.67 ± 0.45 L/min, respectively. At the same target volume but with a Hg of 5.7 g/dL, P was −16 ± 12 mmHg versus −19 ± 12 mmHg and Q was 4 ± 0.1 versus 4 ± 0.4 L/min for the wall-less cannula and control respectively. Likewise, P and Q values with water were −1 mmHg versus −0.67 ± 0.58 mmHg and 4.17 ± 0.45 L/min versus 4.08 ± 0.47 L/min for the wall-less and control respectively. Conclusion: Walls-less cannulaAim: This study was designed to quantify the influence of blood as test medium compared to water in cannula bench performance assessment. Methods: An in vitro circuit was set-up with silicone tubing between two reservoirs. The test medium was pumped from the lower reservoir by centrifugal pump to the upper reservoir. The test-cannula was inserted in a silicone tube connected between the lower reservoir and the centrifugal pump. Flow rate and pump inlet-pressure were measured for wall-less versus thin-wall cannula using a centrifugal pump in a dynamic bench-test for an afterload of 40-60 mmHg using two media: blood 10 g/dL and 5.6 g/dL and water 0 g/dL. Results: The wall-less cannula showed significantly higher flows rates as compared to the thin-wall cannula (control), with both hemoglobin concentrations and water. Indeed, for a target volume of 200-250 mL of blood (Hg 10 g/dL) in the upper reservoir, the cannula outlet pressure (P) was −14 ± 14 mmHg versus −18 ± 11 mmHg for the wall-less and control respectively; the cannula outlet flow rate (Q) was 3.91 ± 0.41 versus 3.67 ± 0.45 L/min, respectively. At the same target volume but with a Hg of 5.7 g/dL, P was −16 ± 12 mmHg versus −19 ± 12 mmHg and Q was 4 ± 0.1 versus 4 ± 0.4 L/min for the wall-less cannula and control respectively. Likewise, P and Q values with water were −1 mmHg versus −0.67 ± 0.58 mmHg and 4.17 ± 0.45 L/min versus 4.08 ± 0.47 L/min for the wall-less and control respectively. Conclusion: Walls-less cannula showed 5.6% less pump inlet-pressure differences calculated between blood and water, as compared to that of thin-wall cannula (−21 times). Flow differences were 6% and 10% for the walls-less and thin-wall cannula respectively. We conclude that testing the cannula performance with water is a good scenario and can overestimate the flow by a 10%. However, superiority for wall-less is preserved with both water and blood. … (more)
- Is Part Of:
- Perfusion. Volume 35:Number 5(2020)
- Journal:
- Perfusion
- Issue:
- Volume 35:Number 5(2020)
- Issue Display:
- Volume 35, Issue 5 (2020)
- Year:
- 2020
- Volume:
- 35
- Issue:
- 5
- Issue Sort Value:
- 2020-0035-0005-0000
- Page Start:
- 393
- Page End:
- 396
- Publication Date:
- 2020-07
- Subjects:
- blood -- hemoglobin -- water -- viscosity -- flow -- pressure -- wall-less cannula
Perfusion (Physiology) -- Periodicals
Blood -- Circulation, Artificial -- Periodicals
Heart -- Surgery -- Periodicals
Extracorporeal Circulation -- Periodicals
Perfusion -- Periodicals
Circulation extracorporelle -- Périodiques
Perfusion -- Périodiques
617.41 - Journal URLs:
- http://prf.sagepub.com/ ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0267659119885514 ↗
- Languages:
- English
- ISSNs:
- 0267-6591
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13110.xml