Impact of patient positioning uncertainty in noncoplanar intracranial stereotactic radiotherapy. Issue 2 (20th January 2020)
- Record Type:
- Journal Article
- Title:
- Impact of patient positioning uncertainty in noncoplanar intracranial stereotactic radiotherapy. Issue 2 (20th January 2020)
- Main Title:
- Impact of patient positioning uncertainty in noncoplanar intracranial stereotactic radiotherapy
- Authors:
- Tanaka, Yoshihiro
Oita, Masataka
Inomata, Shinichiro
Fuse, Toshiaki
Akino, Yuichi
Shimomura, Kohei - Abstract:
- Abstract: The aim of this study is to evaluate the patient positioning uncertainty in noncoplanar stereotactic radiosurgery or stereotactic radiotherapy (SRS/SRT) for intracranial lesions with the frameless 6D ExacTrac system. In all, 28 patients treated with SRS/SRT of 70 treatment plans at our institution were evaluated in this study. Two X‐ray images with the frameless 6D ExacTrac system were first acquired to correct (XC) and verify (XV) the patient position at a couch angle of 0º. Subsequently, the XC and XV images were also acquired at each planned couch angle for using noncoplanar beams to detect position errors caused by rotating a couch. The translational XC and XV shift values at each couch angle were calculated for each plan. The percentages of the translational XC shift values within 1.0 mm for each planned couch angle for using noncoplanar beams were 77.86%, 72.26%, and 98.47% for the lateral, longitudinal, and vertical directions, respectively. Those within 2.0 mm were 98.22%, 97.96%, and 99.75% for the lateral, longitudinal, and vertical directions, respectively. The maximum absolute values of the translational XC shifts among all planned couch angles for using noncoplanar beams were 2.69, 2.45, and 2.17 mm for the lateral, longitudinal, and vertical directions, respectively. The overall absolute values of the translational XV shifts were less than 1.0 mm for all directions except for one case in the longitudinal direction. The patient position errors wereAbstract: The aim of this study is to evaluate the patient positioning uncertainty in noncoplanar stereotactic radiosurgery or stereotactic radiotherapy (SRS/SRT) for intracranial lesions with the frameless 6D ExacTrac system. In all, 28 patients treated with SRS/SRT of 70 treatment plans at our institution were evaluated in this study. Two X‐ray images with the frameless 6D ExacTrac system were first acquired to correct (XC) and verify (XV) the patient position at a couch angle of 0º. Subsequently, the XC and XV images were also acquired at each planned couch angle for using noncoplanar beams to detect position errors caused by rotating a couch. The translational XC and XV shift values at each couch angle were calculated for each plan. The percentages of the translational XC shift values within 1.0 mm for each planned couch angle for using noncoplanar beams were 77.86%, 72.26%, and 98.47% for the lateral, longitudinal, and vertical directions, respectively. Those within 2.0 mm were 98.22%, 97.96%, and 99.75% for the lateral, longitudinal, and vertical directions, respectively. The maximum absolute values of the translational XC shifts among all planned couch angles for using noncoplanar beams were 2.69, 2.45, and 2.17 mm for the lateral, longitudinal, and vertical directions, respectively. The overall absolute values of the translational XV shifts were less than 1.0 mm for all directions except for one case in the longitudinal direction. The patient position errors were detected after couch rotation for using noncoplanar beams, and they exceeded a planning target volume (PTV) margin of 1.0–2.0 mm used commonly in SRS/SRT treatment. These errors need to be corrected at each planned couch angle, or the PTV margin should be enlarged. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 21:Issue 2(2020)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 21:Issue 2(2020)
- Issue Display:
- Volume 21, Issue 2 (2020)
- Year:
- 2020
- Volume:
- 21
- Issue:
- 2
- Issue Sort Value:
- 2020-0021-0002-0000
- Page Start:
- 89
- Page End:
- 97
- Publication Date:
- 2020-01-20
- Subjects:
- IGRT -- noncoplanar radiotherapy -- patient positioning uncertainty -- SRS -- SRT
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12820 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13074.xml