Dosimetric evaluation of incorporating the revised V4.0 calibration protocol for breast intraoperative radiotherapy with the INTRABEAM system. Issue 2 (10th February 2020)
- Record Type:
- Journal Article
- Title:
- Dosimetric evaluation of incorporating the revised V4.0 calibration protocol for breast intraoperative radiotherapy with the INTRABEAM system. Issue 2 (10th February 2020)
- Main Title:
- Dosimetric evaluation of incorporating the revised V4.0 calibration protocol for breast intraoperative radiotherapy with the INTRABEAM system
- Authors:
- Shaikh, Mubin Y.
Burmeister, Jay
Scott, Robin
Kumaraswamy, Lalith K.
Nalichowski, Adrian
Joiner, Michael C. - Abstract:
- Abstract: In breast‐targeted intraoperative radiotherapy (TARGIT) clinical trials (TARGIT‐B, TARGIT‐E, TARGIT‐US), a single fraction of radiation is delivered to the tumor bed during surgery with 1.5‐ to 5.0‐cm diameter spherical applicators and an INTRABEAM x‐ray source (XRS). This factory‐calibrated XRS is characterized by two depth‐dose curves (DDCs) named "TARGIT" and "V4.0." Presently, the TARGIT DDC is used to treat patients enrolled in clinical trials; however, the V4.0 DDC is shown to better represent the delivered dose. Therefore, we reevaluate the delivered prescriptions under the TARGIT protocols using the V4.0 DDC. A 20‐Gy dose was prescribed to the surface of the spherical applicator, and the TARGIT DDC was used to calculate the treatment time. For a constant treatment time, the V4.0 DDC was used to recalculate the dosimetry to evaluate differences in dose rate, dose, and equivalent dose in 2‐Gy fractions (EQD2) for an α/β = 3.5 Gy (endpoint of locoregional relapse). At the surface of the tumor bed (i.e., spherical applicator surface), the calculations using the V4.0 DDC predicted increased values for dose rate (43–16%), dose (28.6–23.2 Gy), and EQD2 (95–31%) for the 1.5‐ to 5.0‐cm diameter spherical applicator sizes, respectively. In general, dosimetric differences are greatest for the 1.5‐cm diameter spherical applicator. The results from this study can be interpreted as a reevaluation of dosimetry or the dangers of underdosage, which can occur if the V4.0 DDCAbstract: In breast‐targeted intraoperative radiotherapy (TARGIT) clinical trials (TARGIT‐B, TARGIT‐E, TARGIT‐US), a single fraction of radiation is delivered to the tumor bed during surgery with 1.5‐ to 5.0‐cm diameter spherical applicators and an INTRABEAM x‐ray source (XRS). This factory‐calibrated XRS is characterized by two depth‐dose curves (DDCs) named "TARGIT" and "V4.0." Presently, the TARGIT DDC is used to treat patients enrolled in clinical trials; however, the V4.0 DDC is shown to better represent the delivered dose. Therefore, we reevaluate the delivered prescriptions under the TARGIT protocols using the V4.0 DDC. A 20‐Gy dose was prescribed to the surface of the spherical applicator, and the TARGIT DDC was used to calculate the treatment time. For a constant treatment time, the V4.0 DDC was used to recalculate the dosimetry to evaluate differences in dose rate, dose, and equivalent dose in 2‐Gy fractions (EQD2) for an α/β = 3.5 Gy (endpoint of locoregional relapse). At the surface of the tumor bed (i.e., spherical applicator surface), the calculations using the V4.0 DDC predicted increased values for dose rate (43–16%), dose (28.6–23.2 Gy), and EQD2 (95–31%) for the 1.5‐ to 5.0‐cm diameter spherical applicator sizes, respectively. In general, dosimetric differences are greatest for the 1.5‐cm diameter spherical applicator. The results from this study can be interpreted as a reevaluation of dosimetry or the dangers of underdosage, which can occur if the V4.0 DDC is inadvertently used for TARGIT clinical trial patients. Because the INTRABEAM system is used in TARGIT clinical trials, accurate knowledge about absorbed dose is essential for making meaningful comparisons between radiation treatment modalities, and reproducible treatment delivery is imperative. The results of this study shed light on these concerns. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 21:Issue 2(2020)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 21:Issue 2(2020)
- Issue Display:
- Volume 21, Issue 2 (2020)
- Year:
- 2020
- Volume:
- 21
- Issue:
- 2
- Issue Sort Value:
- 2020-0021-0002-0000
- Page Start:
- 50
- Page End:
- 59
- Publication Date:
- 2020-02-10
- Subjects:
- INTRABEAM -- breast intraoperative radiotherapy -- linear‐quadratic model -- TARGIT -- V4.0 -- spherical applicators
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12807 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13074.xml