Improvement of image quality applying iterative scatter correction for grid-less skeletal radiography in trauma room setting. (June 2020)
- Record Type:
- Journal Article
- Title:
- Improvement of image quality applying iterative scatter correction for grid-less skeletal radiography in trauma room setting. (June 2020)
- Main Title:
- Improvement of image quality applying iterative scatter correction for grid-less skeletal radiography in trauma room setting
- Authors:
- Lisson, Christoph G
Lisson, Catharina S
Vogele, Daniel
Strauss, Beatrice
Schuetze, Konrad
Cintean, Raffael
Beer, Meinrad
Schmidt, Stefan A - Abstract:
- Background: Iterative reconstruction is well established for CT. Plain radiography also takes advantage of iterative algorithms to reduce scatter radiation and improve image quality. First applications have been described for bedside chest X-ray. A recent experimental approach also provided proof of principle for skeletal imaging. Purpose: To examine clinical applicability of iterative scatter correction for skeletal imaging in the trauma setting. Material and Methods: In this retrospective single-center study, 209 grid-less radiographs were routinely acquired in the trauma room for 12 months, with imaging of the chest (n = 31), knee (n = 111), pelvis (n = 14), shoulder (n = 24), and other regions close to the trunk (n = 29). Radiographs were postprocessed with iterative scatter correction, doubling the number of images. The radiographs were then independently evaluated by three radiologists and three surgeons. A five-step rating scale and visual grading characteristics analysis were used. The area under the VGC curve (AUCVGC ) quantified differences in image quality. Results: Images with iterative scatter correction were generally rated significantly better (AUCVGC = 0.59, P < 0.01). This included both radiologists (AUCVGC = 0.61, P < 0.01) and surgeons (AUCVGC = 0.56, P < 0.01). The image-improving effect was significant for all body regions; in detail: chest (AUCVGC = 0.64, P < 0.01), knee (AUCVGC = 0.61, P < 0.01), pelvis (AUCVGC = 0.60, P = 0.01), shoulderBackground: Iterative reconstruction is well established for CT. Plain radiography also takes advantage of iterative algorithms to reduce scatter radiation and improve image quality. First applications have been described for bedside chest X-ray. A recent experimental approach also provided proof of principle for skeletal imaging. Purpose: To examine clinical applicability of iterative scatter correction for skeletal imaging in the trauma setting. Material and Methods: In this retrospective single-center study, 209 grid-less radiographs were routinely acquired in the trauma room for 12 months, with imaging of the chest (n = 31), knee (n = 111), pelvis (n = 14), shoulder (n = 24), and other regions close to the trunk (n = 29). Radiographs were postprocessed with iterative scatter correction, doubling the number of images. The radiographs were then independently evaluated by three radiologists and three surgeons. A five-step rating scale and visual grading characteristics analysis were used. The area under the VGC curve (AUCVGC ) quantified differences in image quality. Results: Images with iterative scatter correction were generally rated significantly better (AUCVGC = 0.59, P < 0.01). This included both radiologists (AUCVGC = 0.61, P < 0.01) and surgeons (AUCVGC = 0.56, P < 0.01). The image-improving effect was significant for all body regions; in detail: chest (AUCVGC = 0.64, P < 0.01), knee (AUCVGC = 0.61, P < 0.01), pelvis (AUCVGC = 0.60, P = 0.01), shoulder (AUCVGC = 0.59, P = 0.02), and others close to the trunk (AUCVGC = 0.59, P < 0.01). Conclusion: Iterative scatter correction improves the image quality of grid-less skeletal radiography in the clinical setting for a wide range of body regions. Therefore, iterative scatter correction may be the future method of choice for free exposure imaging when an anti-scatter grid is omitted due to high risk of tube-detector misalignment. … (more)
- Is Part Of:
- Acta radiologica. Volume 61:Number 6(2020)
- Journal:
- Acta radiologica
- Issue:
- Volume 61:Number 6(2020)
- Issue Display:
- Volume 61, Issue 6 (2020)
- Year:
- 2020
- Volume:
- 61
- Issue:
- 6
- Issue Sort Value:
- 2020-0061-0006-0000
- Page Start:
- 768
- Page End:
- 775
- Publication Date:
- 2020-06
- Subjects:
- Skeletal -- appendicular -- axial -- radiation safety -- digital radiography -- technology assessments
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1177/0284185119878348 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13074.xml