Five-Year Two-Center Retrospective Comparison of Central Laboratory Glucose to GEM 4000 and ABL 800 Blood Glucose: Demonstrating the (In)adequacy of Blood Gas Glucose. (May 2020)
- Record Type:
- Journal Article
- Title:
- Five-Year Two-Center Retrospective Comparison of Central Laboratory Glucose to GEM 4000 and ABL 800 Blood Glucose: Demonstrating the (In)adequacy of Blood Gas Glucose. (May 2020)
- Main Title:
- Five-Year Two-Center Retrospective Comparison of Central Laboratory Glucose to GEM 4000 and ABL 800 Blood Glucose: Demonstrating the (In)adequacy of Blood Gas Glucose
- Authors:
- Cembrowski, George
Jung, Joanna
Mei, Junyi
Xu, Eric
Curic, Tihomir
Gibney, RT Noel
Jacka, Michael
Sadrzadeh, Hossein - Abstract:
- Purpose: To evaluate the glucose assays of two blood gas analyzers (BGAs) in intensive care unit (ICU) patients by comparing ICU BGA glucoses to central laboratory (CL) glucoses of almost simultaneously drawn specimens. Methods: Data repositories provided five years of ICU BGA glucoses and contemporaneously drawn CL glucoses from a Calgary, Alberta ICU equipped with IL GEM 4000 and CL Roche Cobas 8000-C702, and an Edmonton, Alberta ICU equipped with Radiometer ABL 800 and CL Beckman-Coulter DxC. Blood glucose analyzer and CL glucose differences were evaluated if they were both drawn either within ±15 or ±5 minutes. Glucose differences were assessed graphically and quantitatively with simple run charts and the surveillance error grid (SEG) and quantitatively with the 2016 Food and Drug Administration guidance document, with ISO 15197 and SEG statistical summaries. As the GEM glucose exhibits diurnal variation, CL-arterial blood gas (ABG) differences were evaluated according to time of day. Results: Compared to the GEM glucoses measured between 0200 and 0800, the run charts of (GEM-CL) glucose demonstrate significant outliers between 0800 and 0200 which are identified as moderate to severe clinical outliers by SEG analysis ( P < .002 and P < .0005 for 5- and 15-minute intervals). Over the entire 24-hour period, the rates of moderate to severe glucose clinical outliers are 3.5/1000 (GEM) and 0.6/1000 glucoses (ABL), respectively, using the 15-minute interval ( P < .0001).Purpose: To evaluate the glucose assays of two blood gas analyzers (BGAs) in intensive care unit (ICU) patients by comparing ICU BGA glucoses to central laboratory (CL) glucoses of almost simultaneously drawn specimens. Methods: Data repositories provided five years of ICU BGA glucoses and contemporaneously drawn CL glucoses from a Calgary, Alberta ICU equipped with IL GEM 4000 and CL Roche Cobas 8000-C702, and an Edmonton, Alberta ICU equipped with Radiometer ABL 800 and CL Beckman-Coulter DxC. Blood glucose analyzer and CL glucose differences were evaluated if they were both drawn either within ±15 or ±5 minutes. Glucose differences were assessed graphically and quantitatively with simple run charts and the surveillance error grid (SEG) and quantitatively with the 2016 Food and Drug Administration guidance document, with ISO 15197 and SEG statistical summaries. As the GEM glucose exhibits diurnal variation, CL-arterial blood gas (ABG) differences were evaluated according to time of day. Results: Compared to the GEM glucoses measured between 0200 and 0800, the run charts of (GEM-CL) glucose demonstrate significant outliers between 0800 and 0200 which are identified as moderate to severe clinical outliers by SEG analysis ( P < .002 and P < .0005 for 5- and 15-minute intervals). Over the entire 24-hour period, the rates of moderate to severe glucose clinical outliers are 3.5/1000 (GEM) and 0.6/1000 glucoses (ABL), respectively, using the 15-minute interval ( P < .0001). Discussion: The GEM ABG glucose is associated with a higher frequency of moderate to severe glucose clinical outliers, especially between 0800 and 0200, increased CL testing and higher average patient glucoses. … (more)
- Is Part Of:
- Journal of diabetes science and technology. Volume 14:Number 3(2020:May)
- Journal:
- Journal of diabetes science and technology
- Issue:
- Volume 14:Number 3(2020:May)
- Issue Display:
- Volume 14, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 14
- Issue:
- 3
- Issue Sort Value:
- 2020-0014-0003-0000
- Page Start:
- 535
- Page End:
- 545
- Publication Date:
- 2020-05
- Subjects:
- blood gas analyzer -- glucose accuracy -- GEM 4000 -- ICU -- Radiometer ABL 800 -- surveillance error grid
Diabetes -- Periodicals
Medical technology -- Periodicals
Diabetes Mellitus -- Periodicals
616.462005 - Journal URLs:
- http://ejournals.ebsco.com/direct.asp?JournalID=712321 ↗
http://www.jodsat.org/about.html ↗
http://online.sagepub.com/ ↗ - DOI:
- 10.1177/1932296819883260 ↗
- Languages:
- English
- ISSNs:
- 1932-2968
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13059.xml