- A method to correct background phase offset for phase‐contrast MRI in the presence of steady flow and spatial wrap‐around artifact
- Authors:
- Pruitt, Aaron A.
Jin, Ning
Liu, Yingmin
Simonetti, Orlando P.
Ahmad, Rizwan - Abstract:
- Abstract : Purpose: Background phase offsets in phase‐contrast MRI are often corrected using polynomial regression; however, correction performance degrades when temporally invariant outliers such as steady flow or spatial wrap‐around artifact are present. We describe and validate an iterative method called automatic rejection of temporally invariant outliers (ARTO), which excludes these outliers from the fitting process. Methods: The ARTO method iteratively removes pixels with large polynomial regression errors analyzed by a Gaussian mixture model fitting of the residual distribution. A total of 150 trials of a simulated phantom (75 with wrap‐around artifact) and 125 phase‐contrast MRI cines from 22 healthy subjects (48 with wrap‐around artifact) were used for validation. Background phase offsets were corrected using second‐order weighted regularized least squares (WRLS) with and without ARTO. Flow volumes after WRLS and WRLS+ARTO corrections were compared with the known truth (phantom) and stationary phantom reference (in vivo) using Bland‐Altman analysis. The ratio between the pulmonary flow and the systemic flow was also computed in a subset of 6 subjects. Results: In the simulated phantom, compared with WRLS and no correction, correction with WRLS+ARTO produced superior agreement in volumetric flow quantification with the known truth. In vivo, WRLS+ARTO also produced superior agreement with stationary phantom‐corrected volumetric flow compared with WRLS and noAbstract : Purpose: Background phase offsets in phase‐contrast MRI are often corrected using polynomial regression; however, correction performance degrades when temporally invariant outliers such as steady flow or spatial wrap‐around artifact are present. We describe and validate an iterative method called automatic rejection of temporally invariant outliers (ARTO), which excludes these outliers from the fitting process. Methods: The ARTO method iteratively removes pixels with large polynomial regression errors analyzed by a Gaussian mixture model fitting of the residual distribution. A total of 150 trials of a simulated phantom (75 with wrap‐around artifact) and 125 phase‐contrast MRI cines from 22 healthy subjects (48 with wrap‐around artifact) were used for validation. Background phase offsets were corrected using second‐order weighted regularized least squares (WRLS) with and without ARTO. Flow volumes after WRLS and WRLS+ARTO corrections were compared with the known truth (phantom) and stationary phantom reference (in vivo) using Bland‐Altman analysis. The ratio between the pulmonary flow and the systemic flow was also computed in a subset of 6 subjects. Results: In the simulated phantom, compared with WRLS and no correction, correction with WRLS+ARTO produced superior agreement in volumetric flow quantification with the known truth. In vivo, WRLS+ARTO also produced superior agreement with stationary phantom‐corrected volumetric flow compared with WRLS and no correction. In data sets with wrap‐around artifact, WRLS produced significantly larger variance in the pulmonary flow and systemic flow ratio than stationary phantom correction ( P = .0008). Conclusion: The proposed method provides automatic exclusion of temporally invariant outliers and produces flow quantification results comparable to stationary phantom correction. … (more)
- Is Part Of:
- Magnetic resonance in medicine. Volume 81:Issue 4(2019)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 81:Issue 4(2019)
- Issue Display:
- Volume 81, Issue 4 (2019)
- Year:
- 2019
- Volume:
- 81
- Issue:
- 4
- Issue Sort Value:
- 2019-0081-0004-0000
- Page Start:
- 2424
- Page End:
- 2438
- Publication Date:
- 2018-11-15
- Subjects:
- background phase -- background phase correction -- cardiovascular MRI -- eddy currents -- flow quantification -- phase contrast
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.27572 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 13049.xml